The calibration output (3 V
To remove the battery cover, use a coin to turn the slot on the back of the

unit counter-clockwise
Select the DC coupling for the channel connected to the probe

Run an Autoset

Adjust the sensitivity and vertical offset of the channel so that the signal fills

the screen
Adjust the time base to 200 µs to view a signal period on the screen

Turn the BNC base of the probe in to access the probe adjustment screw:
14
, 1 kHz) for the probe is under the battery cover.
pp
Connect the probe to be adjusted to the calibration output
under the battery housing cover, as shown to the left.
to carry out pre-setting
In this example, the probe is overcompensated, so an
overshoot occurs.
Turn the screw in either direction until the signal is
horizontal and looks like the screen to the left.
Your probe is now calibrated.
Turn the BNC base again to close access to the
adjustment screw and replace the battery cover.