APPENDIX C
SUMMARY OF PARAMETER RANGES / ACCURACIES
SUMMARY OF PARAMETER RANGES / ACCURACIES
SUMMARY OF PARAMETER RANGES / ACCURACIES
SUMMARY OF PARAMETER RANGES / ACCURACIES
PARAMETER
RANGE
___________
__________
Rate
5- 150bpm
Tidal Volume
5- 2500ml
Flow Rate
1- 120 lpm
Peak Pres
0- 120 cmH
PEEP Pres
0- 35 cmH
Pressure Trigger
-10 to -0.2cm
Press Supp.
0- 50 cmH
SIMV Rate
.6- 50bpm
O
sensor
18- 100%
2
EXHTV
50– 3200 ml
EXHMV
0-45 L
Accurate from 100-2500 ml
1
Accurate from 1-120 lpm ±10% of setting or 1 lpm, whichever is greater. With entrainment on, flow
2
accuracy from 5 - 100 LPM is ±10% or 1 LPM, whichever is greater, and with a flow of 100 LPM and over,
flow accuracy is within 15% of displayed flow. No accuracy is claimed below 5 LPM.
Readings below 100 ml may be considered relative rather than absolute values.
3
RESOLUTION OF DISPLAY
_________________________
.1 bpm below 10; 1 bpm above 10
1ml below 200; 10 ml above 200
.1 lpm below 10; 1 lpm above 10
O
1 cmH
O
2
2
O
1 cmH
O (above baseline)
2
2
.1 cmH
O below 3; 1cm above 3
2
O
1 cmH
O (above baseline)
2
2
Same as Rate
1%
1 ml below 200; 10 ml above 200
0.1 L below 10 L, 1 L above 10 L
ACCURACY
________________
±10%
±10%
1
±10%
2
±3% FS
±3cm H
O
2
±1cm H
O
2
±3cm H
O
2
±10%
±3% FS
±15%
3
±15%
86
86
86
86