15. Error Messages and Fault Clearance
Messages are displayed in plaintext and are self-explanatory.
Error Message /
Error
"Dry Strip"
"Wrong Strip"
"Wrong Position" The strip hasn't been pushed
"No Paper"
"Battery Low"
Instrument
doesn't start
In case an error cannot be cleared using the instructions above, please contact
the Henry Schein Pro Repair Service (see 17. Service, page 41).
Manual Urispec
Cause
The test strip wasn't dipped
completely
A wrong test strip has been
detected (wrong type)
into the strip retainer far
enough
Paper roll empty or printer flap
open
Batteries are low
Power supply not installed or
defect
Plus V2.01 HS USA / 04-19
®
User manual Urispec
Solution
Repeat measurement with
a new strip
Make sure you selected
the correct strip (see 5.1.
Select strip, page 19)
New measurement, place
strip in right position
Replace paper and close
printer flap
Exchange batteries or use
power pack
Check whether all connec-
tions are plugged in and
whether the power socket
is functioning