Have a replacement product ready for each application or plan for an alternative surgical
technique.

3.10 Observing ambient conditions

If the device is stored, transported, operated or reprocessed under unsuitable conditions, patients,
users or third parties may be injured and the device can be damaged.
Observe the ambient conditions listed in the instructions for use and reprocessing.

3.11 Risk of explosion and fire

The product can generate sparks, which cause combustible or flammable gases and liquids to ignite
or explode. This may cause injuries to patients, users, and third parties.
When using explosive narcotic gases: Operate the product outside of the hazard zone.
Do not use the product in the presence of flammable anesthetics.
The product must not be operated in oxygenated environments.
Only connect or disconnect the power plug to or from the power supply outside explosive
atmospheres.

3.12 Image failure during HF application

If the product is used together with a high frequency surgical unit, accidental contact between the
active electrode and the metal parts of the endoscope can cause the image to fail.
Make sure that the working part of the active electrode is in the user's field of view.
Avoid contact between the active electrode and metal parts of the endoscope.
If the image has failed, proceed as follows:
Stop use immediately.
If the image has not reappeared after 4 seconds, switch the camera control unit off and on again.
Instructions for use • TELE PACK + • CQH848_EN_V2.0_08-2022_IFU_CE-MDR