  • Page 1 0197 CARDIMAX FX-7102 ELECTROCARDIOGRAPH OPERATION MANUAL • Before using this device, read this operation manual. • Store this manual in a safe location near the unit for future reference.
  • Page 2 Copyright 2004 Fukuda Denshi Co., Ltd. No part of this document may be copied or transmitted in any form without the prior written permission of Fukuda Denshi Co., Ltd. Printed in China Note: Only qualified physicians or persons under instructions from physicians are...
  • Page 3 This manual is organized into 7 chapters, with 4 appendices, glossary, and index. Chapter 1, "General Information" deals with a general introduction to the Cardimax FX- 7102. This chapter contains illustrations of the Cardimax FX-7102 so you can become familiar with the equipment before beginning to use.
  • Page 4 Keys The step-by-step operation procedures in this manual are shown with screens and text. When you operate the Cardimax FX-7102, you use the Operation Panel keys. In this manual, the keys are indicated in the following manner. Example of an Operation Panel key: 【START/STOP】...
  • Page 5 Special Notice to User Malfunction caused by EMC The Cardimax FX-7102 complies with the safety standards IEC60601-2-25 (1999) and IEC60601-1-2 (2001). However, if there is a strong electromagnetic wave nearby that exceeds the prescribed limit, the device could malfunction. If this happens, provide the required countermeasures.
  • Page 6 Equipment classification The Cardimax FX-7102 falls under the following equipment classifications: 1) Protection against electrical shock Class II, Internally powered 2) Type against electrical shock Applied part: Type CF 3) Protection against liquids entering the device General equipment standards 4) Operating safety in the presence of air-inflammable anesthetic gases or...
  • Page 7 Safety Precautions Before using the Cardimax FX-7102, be sure to read this section to ensure safety and • correct operation. Be sure to follow the precautions given below. These are important matters related to • safety. Safety notifications The following three safety notices alert the user to the level of hazard and damage that may result from not following the instructions or from misuse of the equipment.
  • Page 8 Precautions on measurement program CAUTION The equipment may measure the ECG waveform incorrectly due to external noise (such as a drift of baseline, electromyogram, AC interference, etc.). Another matter which requires careful consideration is that tiny R waves in V1 and V2 may be overlooked and the S wave could be interpreted as Q wave.
  • Page 9 DANGER • Be sure to use the specified battery pack (8PH-4/3A3700-H- J18). • Be sure to charge the battery using the Cardimax FX-7102. Charging the battery using other equipment could lead to a hazard. • Do not disassemble or mutilate the battery. You could burn yourself.
  • Page 10 Defibrillation protection WARNING The FX-7102 may remain connected to the patient during defibrillation only if the proper patient cable provided with defibrillator protection is being used. The patient cable (CP- 104T, CP-104L or CP-104J) is provided with defibrillation protection. The FX-7102 will not be damaged when these cables are used, even if the defibrillator paddles come into contact with the ECG electrodes.
  • Page 11 Preventive maintenance CAUTION The purpose of preventive maintenance is to assure that the device is always in a safe operating condition and to prevent possible future problems. Preventive maintenance should be performed at least once during each 12-month period. Preventive maintenance consists of a thorough inspection of the device, all connections, and all attached cables for signs of physical damage.
  • Page 12 Such damage to the equipment could cause an electric shock or fire. If you unintentionally drop the FX-7102, contact your local Fukuda Denshi sales and service representative. • Do not subject the liquid crystal display to a strong impact. The device can be damaged by unwarranted force.
  • Page 13 Equipment Symbols ________________________________________________ The following symbols are printed on the Cardimax FX-7102. Attention: Read the Operation Manual. Type CF equipment Potential equalization terminal AC power supply LAN port CONTRAST volume CONTRAST...
  • Page 14 Guidance for Electromagnetic Compatibility (EMC) ________________________ Details about the electromagnetic compatibility (EMC) of the Cardimax FX-7102 are given below. Before using the FX-7102, be sure to read and understand the following information. Compliance with electromagnetic emissions The Cardimax FX-7102 is intended for use in the electromagnetic environment specified in the following table.
  • Page 15 Compliance with electromagnetic immunity 1 The Cardimax FX-7102 is intended for use in the electromagnetic environment specified in the following table. Before using the FX-7102, check the environment of the site where it is used. Immunity test IEC 60601 test...
  • Page 16 Compliance with electromagnetic immunity 2 The Cardimax FX-7102 is intended for use in the electromagnetic environment specified in the following table. Before using the FX-7102, check the environment of the site where it is used. Immunity test IEC 60601 test...
  • Page 17 Recommended separation distance between portable/mobile RF communications equipment and the FX-7102 The Cardimax FX-7102 is intended for use in an electromagnetic environment where RF emission interference is controlled. The electromagnetic interference can be prevented during operation by providing a minimum distance between portable/mobile RF communications equipment (transmitters) and the FX-7102 as recommended below (according to the maximum output power of the transmitter).
  • Page 18 Accuracy of evaluated data __________________________________________ This instrument conforms to safety requirements and achieves basic performance stipulated by safety standards IEC60601-2-25 (1999) and IEC60601-2-51 (2003). The following performance is assured for the tests required in those standards. Automatic ECG measurement 1. How to determine the amplitude of P wave, QRS wave, ST, and T wave Amplitude of P1A and P2A: The first waveform is defined as P1A and the second as P2A, with biphasic property taken into consideration.
  • Page 19 Reading ECG Data from the Built-in Memory................58 Transmitting ECG Data to External Device ................61 Reviewing ECG data........................63 Chapter 6 Setting Up Parameters of the Cardimax FX-7102 ......66 Overview of Parameters ......................66 General Flow of the Parameter Setup Procedure................72 Selecting the Number of Waveform Display Channels..............75 Selecting the Recording Format ....................76...
  • Page 20 Chapter 7 Maintenance and Troubleshooting ..........92 Performing the Self-diagnostics Test...................92 Cardimax FX-7102 Maintenance ....................100 When You cannot Record an Accurate ECG................102 When an Error Message is Displayed ..................103 Appendix A. Specifications ................106 Appendix B. Daily Check List ................109 Appendix C.