WE hope that your departments enjoy with high quality of CAS product. This manual will help you with proper operations and care of the AD series. Please keep it handy for the future references. 1.2. Precaution Make sure that you plug your scale into the proper power outlet.
Service Manual – AD Series REV NO ■ KEY FUNCTIONS KEYS FUNCTIONS Turns the display ON or OFF. Used to correct the ZERO point. Used to enter a TARE weight. Used to remove a TARE weight. When the weight is not stable, display will show the average weight for 4 seconds.
Service Manual – AD Series REV NO 3. Getting Started 3.1. User setup ① While pressing the PRINT key, press the POWER key. The display shows “U SEt”. ② Press the TARE key, the display shows “Pr off”. (If the setting value already exist in the scale, then it would display as the setting.) ③...
Service Manual – AD Series REV NO 4. Calibration Mode 4.1. Genaral Calibration (1) Hold down “Calibration Switch” and “[POWER] key” to enter Calibration mode and then the scale displays “CAL 1” after “onE”. (2) User can move to other menu by using [TARE] key (Next) or [HOLD] key (Previous).
Service Manual – AD Series REV NO 4.2. C4 Setting AD PLUS stores optional settings by using hexadecimal number in C-4 menu. When you entered CAL-4, it will display hexadecimal number on the Total Price Display panel. For example, it displays B0 in C41 setting.
Service Manual – AD Series REV NO 4.2.2. C4-3 Setting (Sale functions) "." dot BIT7 Dot Type "," comma Don't use Use Preset tare BIT6 (AD can’t use) Don't use Use Back light BIT5 (AD can’t use) Don't use Use Head message BIT4 (AD can’t use)
Service Manual – AD Series REV NO (13) Press [ZERO] key, and then “dual “ message blinks.“1” will be shown. The third digit,”1” will blink. It means a dual interval is disable. (0 : disable, 1 : enable)” (14) Input a dual interval enable by using [HOLD] key.
Service Manual – AD Series REV NO 4.5. Mode Setting (C-6) (1) Press [TARE] to display “C-6”. (2) Press [ZERO] key, and then “ HoLd“ or “Print” or “Unit” message (3) Press [TARE] and [HOLD] key at the same time. You can change the Setting Mode.
The "Err 1" occurs when a current zero "Err 1" point has shifted from the last span Please call your CAS dealer. calibration. The "Err 2" is not a real error. Only it "Err 2" prompts return CAL switch to the Please call your CAS dealer.
The "Err 14" means calibration range is “Err 14” Please call your CAS dealer. not correct. 6. Update 6.1. ROM Download Method (1) Connect a RS-232C Cable, between the scale and PC and then excute a O/M Downloader...
Page 15
Service Manual – AD Series REV NO (2) Check a ‘Communication port’ and click the ‘Init. Erase Command’. And then Click the ‘Start Download’, Communication port will be “Ready” status. (3) Push a ‘Open ROM File’ button and then open the ROM File.
Service Manual – AD Series REV NO (4) If click the ‘Start Download’ holding a power ON/OFF key, You will see ROM Downloading display and then ROM download will be finished. 7. Exploded Views & Parts List 7.1. Exploded View...
Service Manual – AD Series REV NO 7.5 Part List 7.5.1. MAIN PCB ASS’Y Part Name Specification Part Number Q'ty Remark 1 PCB-MAIN 6100-PAP-0901-A 6100PAP0901A MAIN PCB 2 ONE MODULE ONE MODULE 6PA0A0000010 U1 ONE MODULE(FULL OPTION) 3 IC(FIP DRIVER)
Service Manual – AD Series REV NO 7.5.3. BODY ASS’Y Part Name Specification Part Number Q'ty Remark TRAY ANGEL(0.9T) 1000A0000120 PLATFORM ANGEL (일반) 샌딩 1100AZ100240 SCREW-MACHINE(PH) M4*6 1502A0004060 SCREW-MACHINE(PH) M4*30 1502A0004300 SCREW-MACHINE(WPH) M4*12 1503A0004120 SCREW-TAPPING(PH)-1 M4*20 1510A0004200 BOLT-WRENCH M6*25-SUS 1530MSU06250...
Page 22
Service Manual – AD Series REV NO 7.5.5. UPPER CASE ASS’Y Part Name Specification Part Number Q'ty Remark SCREW-TAPPING(PH)-1 M3*8 1510A0003080 RIVET @3.2*8 1563A0003080 NAME PLATE AP-M(양면테잎용)ENGLISH 1800APM0000A SPEC PLATE AP,AD(CE) 1810AP000222 K/B SWITCH KNOB 11.4*11.4*6(DB,AD) 2000A000057A LENS ACRYL ø24*10(ANGEL)