Julabo 200F Original Operating Manual

Julabo 200F Original Operating Manual

Refrigerated bath basic units
Hide thumbs Also See for 200F:

Advertisement

Refrigerated bath basic units
Original operating manual
1.950.0170.en.V08
10/2022

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 200F and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Julabo 200F

  • Page 1 Refrigerated bath basic units Original operating manual 1.950.0170.en.V08 10/2022...
  • Page 2 Legal JULABO GmbH Gerhard-Juchheim-Strasse 1 77960 Seelbach / Germany Tel.: +49 7823 51-0 Fax: +49 7823 2491 Info.de@julabo.com www.julabo.com The content of this operating manual is protected by copyright. Information, including texts, images, and other contents may not be reproduced, distributed, transmitted, stored or otherwise used in any form without prior explicit written consent.
  • Page 3: Table Of Contents

    Foreword .................... 5 About this manual ................6 Notes on commissioning and operation ......... 6 Original JULABO spare parts ............6 Accessories ..................6 Warnings ..................7 Symbols used ................. 8 Intended use ..................8 Safety ....................9 Safety instructions ................9 Safety symbols ................
  • Page 4 Device disposal ................32 EC Declaration of Conformity ............33 UK Declaration of Conformity ............40...
  • Page 5: Foreword

    Foreword Congratulations! You have made an excellent choice. JULABO would like to thank you for the trust you have placed in our company and products. This operating manual will help you become acquainted with the use of our units. Read the operating manual carefully. Keep the operating manual handy at...
  • Page 6: About This Manual

    Read the Safety section of this manual before using the equipment for the first time. Notes on commissioning and operation The cooling machine is combined with a suitable JULABO circulator to form a refrigerated circulator. Operation is done entirely via the control elements of the connected circulator.
  • Page 7: Warnings

    About this manual Warnings The manual contains warnings to increase safety when using the device. Warnings must always be observed. A warning sign displayed in signal color precedes the signal word. The signal word, highlighted in color, specifies the severity of the hazard. CAUTION This signal word designates a danger with a low level of risk which, if it not prevented, may result in minor to moderate injuries.
  • Page 8: Symbols Used

    This is a laboratory device intended for temperature control applications of liquid media. It can only be operated in conjunction with a JULABO circulator designed for this purpose. Only use the device if it is in technically perfect condition and only use it in accordance with its intended use.
  • Page 9: Safety

    Touching damaged live parts can cause severe electric shocks and lead to injury or even death. • Have damaged insulation and parts of the electrical system immediately repaired by JULABO service technicians or a qualified specialist workshop • Immediately replace damaged power cords •...
  • Page 10 Refrigerants and their vapors are harmful to health. There is a suffocation risk in enclosed spaces. • Do not touch or inhale refrigerants. • Have damage to the refrigerant cycle repaired only by JULABO service technicians or qualified specialists. • If refrigerant leaks, stop the device immediately and ventilate the room thoroughly.
  • Page 11: Safety Symbols

    Only carry out work described in this operating manual. Switch off the unit and disconnect it from the power supply before carrying out any work. • All other maintenance and repair work may only be carried out by a JULABO service technician or a qualified specialist workshop. Safety symbols There are safety symbols included with the device, which should be attached to the device before initial operation.
  • Page 12: Safety Function

    Safety Safety function Technical safety functions protect the device from damage. The safety functions are not affected by the control circuit. If a safety function trips, all actuators are permanently switched off. The operator is warned by an optical and acoustic alarm on the circulator.
  • Page 13: Product Description

    The cooling machine is controlled by the connected circulator. Possible combinations with circulators The table lists which cooling machine can be combined with which circulator. Cooling Circulators machine CORIO CD CORIO CP DYNEO DD MAGIO MS 200F    201F    300F  ...
  • Page 14: Operating And Functional Elements

    Product description Operating and functional elements The following figure shows the operating and functional elements and their position on the unit. Front (without cover) and rear (e.g. 310F) Recessed grip Bath tank drain opening Drain valve Drain hose (for 310F and 450F) Condenser Recessed grip CAN plug for connection with a circulator...
  • Page 15: 5.3.1 External Inverter

    Product description 5.3.1 External inverter With the 100 V voltage variant of 310F and 450F, the inverter is mounted on the rear side. Rear side of 310F/450F (100 V voltage variant) Inverter CAN plug for connection with a circulator Mains input socket Circulator mains output socket Mains fuse Mains fuse...
  • Page 16: Technical Data

    Product description Technical data Performance specifications measured in accordance with DIN12876. Cooling capacities up to 20°C measured with ethanol; over 20°C with thermal oil unless specified otherwise. Performance specifications apply at an ambient temperature of 20°C. Performance values may differ with other bath fluids. Grouping of the device acc.
  • Page 17 Product description Technical data 200F Performance values Lowest temperature °C Refrigerants R134a Dimensions Dimensions (W x D x H) 23 x 39 x 44 Usable bath opening (W x D) 13 x 15 Bath depth Volumes min. … max. 3.0 … 4.0...
  • Page 18 Product description Mains connection 100 V 115 V 230 V 50/60 Hz 60 Hz 50/60 Hz Nominal current consumption Total current consumption Switzerland Great Britain Mains fuse, resettable Technical data 300F Performance values Lowest temperature °C Refrigerants R134a Dimensions Dimensions (W x D x H) 24 x 42 x 66 Usable bath opening (W x D) 13 x 15...
  • Page 19 Product description Technical data 310F Performance values Lowest temperature °C Refrigerants R449A/R290 Dimensions Dimensions (W x D x H) 23 x 40 x 43 Usable bath opening (W x D) 13 x 15 Bath depth Volumes min. … max. 3.0 … 4.0 Weight Mains connection 100 V...
  • Page 20 Product description Mains connection 100 V 115 V 200-230 V 50/60 Hz 60 Hz 50/60 Hz Nominal current consumption 3 (200 V) 3 (230 V) Total current consumption Switzerland Great Britain Mains fuse, resettable Technical data 600F Performance values Lowest temperature °C Refrigerants R449A, R452A*...
  • Page 21 Product description Technical data 601F Performance values Lowest temperature °C -35/-40* Refrigerants R449A, R452A** Dimensions Dimensions (W x D x H) 33 x 46 x 52 Usable bath opening (W x D) 22 x 15 Bath depth Volumes min. … max. 8.0 …...
  • Page 22 Product description Mains connection 100-230 V 50/60 Hz Nominal current consumption 5 (100 V) 3 (230 V) Total current consumption Mains fuse, resettable at 100 V/ Technical data 900F Performance values Lowest temperature °C -38/-40* Refrigerants R449A Dimensions Dimensions (W x D x H) 39 x 62 x 53 Usable bath opening (W x D) 26 x 35...
  • Page 23 Product description Technical data 1000F Performance values Lowest temperature °C -40/-50* Refrigerants R449A Dimensions Dimensions (W x D x H) 42 x 49 x 51 Usable bath opening (W x D) 18 x 13 Bath depth Volumes min. … max. 5.0 …...
  • Page 24: 5.4.1 Material Of Parts That Come Into Contact With The Medium

    Product description Mains connection 200-230 V 50/60 Hz Nominal current consumption 5 (200 V) 5 (230 V) Total current consumption Switzerland Great Britain Mains fuse, resettable 5.4.1 Material of parts that come into contact with the medium The table lists parts that could come into contact with the bath fluid as well as the material that the parts are made of.
  • Page 25: Transport And Installation

    Transport and installation Transport and installation This section describes how to transport the unit safely. CAUTION Risk of crushing by falling device! A device that is not secured appropriately can fall down during improper transport and cause crushing injuries. • Secure the device against tipping and falling during transport •...
  • Page 26: Maintenance

    Insufficiently dimensioned power cords must not be used. See type plate for mains voltage and current value. We recommend only using original JULABO spare parts. Check safety symbols The following section applies only to equipment working with natural refrigerants and operated in the US or Canada.
  • Page 27 [3] Note (order no.: 3.383.2640): 1. Check the safety symbols on the unit for legibility and completeness. 2. Replace defective or missing safety symbols.  Safety symbols can be reordered from JULABO.  The safety symbols on the unit have been checked.
  • Page 28: Emptying

    Maintenance Emptying The device must be completely drained if it is to be sent in for technical service or is to be properly disposed of. In general, the device should be completely emptied before longer shutdowns or when there is a change to the external application. CAUTION Risk of burns from hot bath fluid! Bath fluid can become very hot during a temperature control process.
  • Page 29: Clean Device

    Maintenance Clean device The outside of the device should be periodically cleaned.  Lint-free cloth  Mild cleaning agent NOTE Damage to the electronics due to water penetration! Ingress of water can damage electronic components of the device and thus lead to failure of the device.
  • Page 30: Device Storage

    Maintenance 3. Use a vacuum cleaner to carefully vacuum the dirt off the condenser.  Make sure that the fins of the condenser are not damaged. 4. Place the venting grid back on.  The condenser is cleaned. Device storage Take your device out of operation if you have not used it for a long time or, for example, it is to be sent to Technical Service for repair.
  • Page 31: Technical Service

    Package the device safely for shipment. Warranty JULABO provides a warranty that the device will function perfectly as long as it is connected and used correctly and as described in the operating manual. The warranty period is one year from the invoice date.
  • Page 32: Disposal

    Disposal Disposal Device disposal When disposing of the device, the applicable country-specific guidelines must be observed. CAUTION Risk of fire with flammable refrigerants! If the device contains a flammable refrigerant, there is a risk of fire if refrigerant leaks. • Do not open the refrigerant circuit •...
  • Page 33 EC Declaration of Conformity EC Declaration of Conformity...
  • Page 34 EC Declaration of Conformity...
  • Page 35 EC Declaration of Conformity...
  • Page 36 EC Declaration of Conformity...
  • Page 37 EC Declaration of Conformity...
  • Page 38 EC Declaration of Conformity...
  • Page 39 EC Declaration of Conformity...
  • Page 40 UK Declaration of Conformity 10 UK Declaration of Conformity UK Office: JULABO UK Ltd., Unit 7, Casterton Road Business Park, Old Great North Road, Little Casterton, Stamford, PE9 4EJ, United Kingdom, Tel.: +44 1733 265892...
  • Page 41 UK Declaration of Conformity...
  • Page 42 UK Declaration of Conformity...
  • Page 43 UK Declaration of Conformity...
  • Page 44 UK Declaration of Conformity...
  • Page 45 UK Declaration of Conformity...

This manual is also suitable for: