YO Loop
Assembly-Level Troubleshooting
INTRODUCTION
The
YO UNLOCKED
error condition can be caused by
a
number of different failures. Figure D-5, YO
Loop Troubleshooting Flow Chart, provides
a
systematic way to identify the malfunctioning assembly.
Also refer to Figure D-4,
Sweep Generator
and
YO
Loop Simplified Block Diagram.
Each
of
the
following
headings refer to
the
Troubleshooting Flow Chart and assumes that the front panel UNLK
indicator
is
on.
Power levels are shown
on Figure D-6.
Sweep Generator and YO Loop Troubleshooting Block
Dia¬
gram.
YO UNLK
-
CW/MANUAL MODE
NOTE:
The
[SHIFT] [M4]
diagnostic can help isolate the cause of an
UNLK
condition. Use this
diagnostic
feature after performing
the troubleshooting
procedure shown below. Refer
to the overall
instrument troubleshooting section
in
the service introduction for details.The instrument
UNLK
diag¬
nostics can be selected by pressing
[SHIFT] [EXT]
on the front panel. This will display the following:
OSC: REF
M/N HET
YO N2
N1
A
YO unlock
condition exists
when the
"YO"
indicator
is
flashing.
In
most cases,
the
error condition
will
occur
in
the
CW/MANUAL
mode, and the troubleshooting
should be done
in
this
mode
first. If
the YO loop
will
not phase-lock
in
CW/MANUAL
at
all
frequencies,
it
will
not function properly
in
the
SWEPT
mode.
To troubleshoot
in
the
CW/MANUAL
mode, perform the following steps.
Refer
to
Figure D-5:
1.
Are Other Loops Locked?
Check
to
see
that the UNLK indicators for the REF,
M/N,
N2, and
N1
loops are
not
flashing.
If
these other
loops
are
unlocked, the
YO
loop will not function correctly. Troubleshoot
the other
loops
first.
If
they
are
all
locked, proceed to question 2.
2.
Is
Front Panel RF Power Correct?
Check that RF
is
present
at the front panel at
a
level
close
to the selected
power.
If
not,
there
is
a
problem
in
the RF path
go to question
14.
If
the
front panel
RF
is
present at the correct power,
proceed
to
question 3.
HP 8340B/41B
YO
Loop
Assembly-Level Troubleshooting