A33J2 Output Check - HP 8340B Manual

Hide thumbs Also See for 8340B:
Table of Contents

Advertisement

A33J2 OUTPUT CHECK
Test Setup
ON THE DUT:
The DUT
should
still
be set to
a
CW
frequency of
4530 MHz
with
its
10
MHz REF
connected to the
spectrum analyzer's
FREQ REFERENCE
input.
Disconnect the power supply from A33TP1 (TUNE).
Connect A33J2 directly to the spectrum analyzer's
input.
ON THE SPECTRUM ANALYZER:
As shown
in
figures on the following page.
Proper Waveform Parameters
As shown
in
views
A
and
B
on the following page,
and within the following
parameters:
Frequency
=
190MHz
Power
=
approximately
0
dBm
±
3
dBm
Harmonics
=
-s
—20 dBc
B-32
M/N
Loop Assembly-Level Troubleshooting