RAUMEDIC MPR2 logO Instructions For Use Manual page 96

Table of Contents

Advertisement

Instructions for Use RAUMEDIC
Analogue outputs
Number of channels:
Transferable input values:
Characteristic features:
Measuring range:
Resolution:
Accuracy after zeroing
*1
Accuracy after zeroing:
Recommended supply voltage:
Maximum supply voltage:
Output frequency:
Additional thermal drift:
*1: Zeroing of the downstream monitor
Interface for the extension of the apparatus
Type:
Application:
PC interface
Type:
Application:
General data
Type designation:
Mode of operation:
Volume of alarm:
Display:
Trend display:
Alarm limits:
Language:
Dimensions:
Mass:
Fuse in battery compartment *2:
Power supply:
Type of battery:
Operating period:
Charger / mains unit:
Class of protection:
IP protection class:
*2: The fuse may be replaced by trained staff only. If the apparatus does not operate after the first
replacement, remove the fuse again and take the apparatus to the after-sales service department for
repair. The fuse can be obtained from the manufacturer, if necessary.
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
MPR2 logO
2
invasive pressure, partial oxygen pressure
5 µV / V / mmHg
- 20 ... + 360 mmHg
< 0.5 mmHg, typical 0.23 mmHg
:
± 1.5 mmHg at 22 °Celsius ambient temperature
± 2.0 mmHg at 10 °Celsius – 40 °Celsius ambient temperature
4 – 6 V DC
10 Vss
100 Hz
max. 0.5 mmHg / K only with a software version lower than
1.06.0032 or 1.06.0032 ST (cf. 4.6.2 Device diagnostics)
RS232 (V24) + power supply
unit service and optional extension of the apparatus
USB
connection of laptop
MPR2 logO
continuous operation
minimum 45 dB(A) typical 60 dB(A), 1 m, in stand holder
black/white LCD 320 x 240 pixel (1/4 VGA), 74 x 97 mm
optional 4 min., 8 min., 20 min., 40 min., 80 min. or 4 h
to be selected by menu
German, other language versions optional
approx. 200 x 69 x 150 mm (W x D x H)
approx. 1.02 kg with battery
type: TR5 / No. 372 / 250V / 1.25 AT / Wickmann
internal battery
NiMH, Panasonic HHR210AH-1Z
> 4 h (charged, new battery pack, all channels occupied, lighting
off, PHB = 1 s, at 22 °Celsius)
Wide range adapter, Art. No. 284007-002
or Art. No. 284027-001
Input: 100-240 V / 50-60 Hz / 400-200 mA,
Output: 12 V / 1.5 A, protection class II, IP 40
-
class of protection II with internal and external power supply
-
type CF IEC 60601-1
-
Directive 93/42/EEC Class II b
IP 21 in operation (sockets to the bottom, maximum inclination
20°)