Instructions for Use RAUMEDIC
If another (second) pressure sensor is connected, the measuring location allocated to P1 is not
available any more in this menu. The same applies corresponding, if you want to connect a third
pressure sensor.
The ICP1 measuring location is available only when the catheter has been connected up to port
P1.
The ICP2 measuring location is available only when the catheter has been connected up to the
P2/T2 port.
If you have determined a measuring location, have adjusted the pressure sensor to zero and
thus have started the measurement, a subsequent modification of the measuring location is not
possible easily. If you want to re-define the measuring location nonetheless, remove the
connector of the connecting cable for the pressure sensor on the MPR2 and reconnect.
Press the menu key; open the
selection
for
the
location.
In the next step, the pressure sensor has to be zeroed. For this purpose, skip to the Zero calibration pull-
down menu. Press the Menu key and select the zeroing procedure recommended for the pressure sensor.
The default setting in this menu is "Zeroing RAUMEDIC catheter". Prior to zeroing, please ensure that the
pressure sensor is subjected to the pressure conditions described in the instructions for use.
Please observe the instructions in the instructions for use of the manufacturer of the pressure
sensor. Faulty zeroing will lead to systematic measuring errors.
Press the Menu key.
Now trigger the zeroing process by pressing the OK key which will be confirmed by two short beeps. The
display will automatically skip to the overview screen with the display of the measured values.
If no other sensors have been connected, the selected measuring location will be displayed.
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
MPR2 logO
Use the arrow keys to select the
measuring
requested
and then acknowledge by pressing
OK.
Select the zeroing procedure and
acknowledge by pressing OK.
Page 38 of 99
Display of the selected measuring
measuring
location,
location.
Display of the selected zeroing
procedure.
zwo-400EN
Rev. 9 2020-09-22