If a new alarm is added, the acoustic alarm of the corresponding priority of the new alarm is added.
The alarm key has no influence on the optical alarm signal (red LED or yellow LED and flashing display on
LCD). As long as the reason for the alarm exists, the optical alarm will be signalled, and the alarm LED
flashes according to the highest priority given.
As to which AUDIO OFF function (AUDIO PAUSING or global AUDIO OFF) is used, is determined by the
setting in the operator settings.
Global AUDIO OFF:
If this key Alarm is pressed for two seconds, all acoustic alarms are globally suppressed until this key is
pressed again or until the MPR2 is switched off. The condition of alarm suppression is permanently
indicated by a flashing symbol in the information area
acoustic alarms. For this reason, the user should be handling this function very knowingly.
The operator settings (cf. Instruction for Use of Software DataView, Device setup MPR 1 - / MPR2 logO
DATALOGGER) can be used to activate a reminder signal. In case of global AUDIO OFF, the reminder
signal releases a beep of 0.5 sec. every 5 minutes.
AUDIO OFF can be used even when the key lock has been switched on. Global AUDIO OFF, however, is
not possible when the key lock has been switched on.
Global AUDIO PAUSING:
If this key Alarm is pressed for two seconds, all acoustic alarms are suppressed for an adjustable interval
or until this key is pressed again or until the MPR2 is switched off. The condition of the alarm suppression
is indicated continuously in the info area by a flashing symbol
lines) and by the count-down timer.
A short depression of the alarm key switches the count-down timer off; a long depression of the alarm
key switches the timer on again. In this case the system cannot produce any acoustic alarms. For this
reason, the user should be handling this function very knowingly.
AUDIO OFF can be used even when the key lock has been switched on. Global AUDIO PAUSING,
however, is not possible when the key lock has been switched on.
A permanently pressed key Alarm
acoustic alarm added will also cause an audible signal which can be switched off only by pressing the alarm
key again.
The following alarms are activated:
Physiological alarms
Technical alarms
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
MPR2 logO
does not mute any new alarms immediately. Thus, every new
Page 70 of 99
. In this case the system cannot produce any
(bell system crossed out by dashed
zwo-400EN
Rev. 9 2020-09-22