RAUMEDIC MPR2 logO Instructions For Use Manual page 47

Table of Contents

Advertisement

Instructions for Use RAUMEDIC
Step 4:
If the pO2 key was pressed in step
1, the setting options for alarms,
signal relaying and control of the
pO2
measuring
intervals
accessed.
The
limiting
values are set identically with the
invasive
pressure;
applies for the signal relaying to a
third-party system. If the Fast key
is pressed, the measuring interval
of 1 sec. will be switched on for
180 sec., independent of the
previous setting, and the window
is closed.
If screen 1 is used to call the adjustment dialogue for pO2 when a PTO catheter has been connected
and an error message is active in the area of the pO2 value, than error message well appear:
"An error has occurred in the O2 measuring system. Adjustments are not possible."
Please observe the special features when setting the measuring interval. The measuring
intervals have influence on the dynamics of the pO2 parameters to be displayed. A short
measuring interval causes a higher consumption of power, thus leading to a reduced mobile
time, than a longer measuring interval. Please keep in mind that the partial oxygen pressure is a
parameter which changes slowly.
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
MPR2 logO
Step 5:
If the Menu key is pressed in step
4, the selection of the pO2
measuring interval is accessed.
are
The pO2 measuring interval can
alarm
be set to 1 sec, 5 sec, 10 sec, 20
sec or 60 sec. In addition, the
the
same
setting can be set to "off", if and
when required. The setting of the
pO2 measuring intervals has a
distinct
achievable mobile time when the
MPR2 is operated on battery pack.
Page 47 of 99
influence
on
the
Step 6:
After acknowledgement of the
selected measuring interval, this is
acknowledged by pressing OK.
The settings will become active
only after this window has been
closed with the OK key, and the
system
has
returned
overview (screen step 1).