Instructions for Use RAUMEDIC
Prior to using the MPR2, the correct condition of the apparatus and any accessories shall have to be
checked. Check the plausibility of the readings before using the apparatus for monitoring. The
apparatus and the accessories must not be damaged or soiled; otherwise the apparatus must not be
used. Do not use the apparatus, if it reveals obvious signs of a wrong function; in this case forward the
unit to the after-sales service of the manufacturer.
In questions of servicing, please always get in touch with the manufacturer. The manufacturer will give
you a list of authorised service partners, if and when required.
In all fastening and erection variants, please make sure that the apparatus cannot drop or cause other
hazards. Please also ensure a safe laying of the cables so that a tipping over of the arrangement is
prevented.
Do not use the apparatus in potentially explosive surroundings. The apparatus shall have to be set up
at a minimum distance of 25 cm to any anaesthesia units and gas filled hoses.
After use the accessories shall have to be cleaned, disinfected or sterilised in keeping with the
instructions for use. If the accessory has been provided with separate instructions for use, the
instructions rendered there shall be applicable.
If the apparatus has been taken from colder into warmer surroundings, it may be used only when the
temperature of the apparatus and of the accessories has adapted to the room temperature; there is no
danger of condensation water forming or any condensation water already developed has completely
dried.
Please also observe the safety instructions in conjunction with the use of sensors and transducers
which have been included to the respective sections of the instructions for use.
No liquids or fluids shall enter the apparatus. If this should happen nonetheless, first remove the mains
adapter from the port. Hand over the apparatus to the after-sales service department for inspection.
Subsequently a safety-related check is necessary.
The MPR2 has been designed in conformity with standard IEC 60601-1. It is a product of class of
protection II with an internal power source and an external mains adapter, and has been allocated to
class IIb (MPG).
If sterile accessories are used, please make sure that they are marked as being sterile.
If the apparatus is subjected to strong electro-magnetic fields, please observe that the pressure display
may vary by up to 2 mmHg. For this reason, avoid using mobile phones, therapeutical microwave
devices or similar units in the vicinity of the apparatus.
The following instructions shall have to be observed for the installation of the system (with laptop and/or
foreign unit – monitor):
Moveable multiple sockets must not lay on the floor.
Additional moveable multiple sockets or extension cables must not be connected up to the system.
Apparatuses, which are not part of the system, must not be connected.
The moveable multiple sockets have to be suitable for the load of the system.
The moveable multiple sockets which are supplied along with the system only may be used for the
supply of the devices of the system.
Observe the instructions for the installer on how to set up the system in an ideal way. Please also
read chapters 3.4, 3.5 and 3.6.
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
MPR2 logO
Page 7 of 99
zwo-400EN
Rev. 9 2020-09-22