Customer menu Table of Content Customer menu ........3 Set up a new customer ........3 Search function: customer, project, customer number, project number ........ 4 Delete customer ..........6 Delete all ............6 Select customers before measurement ..6 Save data in the Readings Menu ..
Customer menu Customer menu The customer menu offers the possibility to create customer folders, to select custom- ers, to display and print already stored measured values or to transfer them. Measurement data can be assigned to different customers and projects in the device. The assign- ment is done via the corresponding customer and project numbers.
Customer menu One after the other, enter the customer • name, the corresponding customer number, the project name and the project number and confirm each with the check mark. Fig. 2: create a new customer file Search function: customer, project, customer number, project number In the customer menu click on the “Find •...
Page 5
Customer menu Enter any known part of the customer • name. You do not need to be case- sensitive. The first matches are now displayed. With the scroll bar you can scroll through the matches and select the desired customers. Fig.
Customer menu Delete customer In the customer menu it is possible to cancel a selected customer. • Click on a customer to select him. • Click on the DELETE CUSTOMER icon. Delete all To delete all customers with the stored data, •...
Customer menu Save data in the Readings Menu If the user accepts a record, the analyzer will en- Readings Icon ter the readings menu automatically. It is also possible to enter the readings menu by clicking on the readings icon in the main menu. Open a customer in the customer menu and •...
NOTE! Position the printer so that there is no obstacle between the infrared interfaces of the printer and the Wöhler M 603. Save measurement data Click on the Save icon to assign all selected •...
Customer menu Menu „Settings“ Device In the submenu “Device” all necessary settings for the device and the measurements can be made. Only 7 of the 14 parameters are displayed in the display view. • Swipe your finger across the display to view additional sub-items.
Page 10
Customer menu If you want to reload the factory default set- • Factory settings tings, tap the Factory Settings icon in the footer. Date Enter the date with day, month and year. Time Enter the time in 24h format with hour and minutes.
Page 11
Customer menu Maintenance instructions If the maintenance note is activated, a note on the recommended date of the next maintenance ap- pears in the main menu. The maintenance date is either 15 months after the date of production and purchase or 12 months after the last calibration. If the maintenance date is exceeded, an exclama- tion mark appears next to the maintenance date.
Page 12
Customer menu Remove marked value: Delete value from the display view. Logo Click on "Change", enter your company name • and confirm with OK. NOTE! During text input, only those letters are taken into account that can be seen in the white input field. Your logo will then appear on all printouts.
Tap “Disconnect” in the same menu to dis- • Fig. 8: Connect Wöhler M 603 to Wöhler connect again. After a short search, the Wöhler M 603 connects to the Wöhler SC 660 units within range. The connected units are displayed under Settings > Smart Connects.
Here you will find important information about the Wöhler M 603 and calibration. Tightness test Before the Wöhler M 603 is used the tightness of the device, of the measuring hoses and of the connectors should be tested. Please, proceed as follows: Switch on the Wöhler M 603.
Customer menu Troubleshooting Error message Possible reason Solution Batteries low Disposable or rechargeable Recharge batteries. batteries are empty. Stabilization not possible. - During the measurement Check connections. The loss of pressure is too the pressure is falling, so high. that no measurement will be possible.
If used properly, the warranty period for the Wöh- ler M 603 will be twelve month from the date of sale. Not covered by the warranty are batteries and defects of the pressure sensor caused by overpressure.
WÖHLER Technik GmbH Wöhler-Platz 1, D-33181 Bad Wünnenberg declares that the product: Product name: Wöhler M 603 Pressure Meter complies with the essential protection requirements laid down in Directive 2014/53/EU of the European Parliament and of the Council on the harmonisa- tion of the laws of the Member States relating to the making available on the market of radio equipment.
Page 18
Gneisenaustr.12 80992 München Tel.: +49 89 1589223-0 Fax: +49 89 1589223-99 sued@woehler.de Czech Republic Wohler USA Inc. Wöhler Bohemia s.r.o. 208 S Main Street Za Naspern 1993 Middleton, MA 01949 393 01 Pelhrimov Tel.: +1 978 750 9876 Tel.: +420 565 349 011 Fax.: +1 978 750 9799...