4-8
T6900 Rev A
Electronic Cycle Tapes/Printouts
Electronic cycle tapes and optional paper printouts record useful information
about each cycle the Processor runs. Cycle tapes include key data:
• cycle name
• date
• time the cycle was started (inflate seal)
• accumulative cycle count
• serial number of the Processor
• "Load" Operator ID, name or codes of the person starting the cycle
• various inputted ID names or codes
• min/max exposure temperatures
• concentration (units of conductivity of use dilution)
• exposure time duration
• fill time
• inlet temperature
• time the cycle was completed
• total cycle time
• "Unload" Operator ID, name or code of the person unloading the Processor
• time the lid was unsealed and opened
• remaining life of the MaxLife water filter (if 5 days or less)
Cycle tapes from aborted/cancelled cycles and alarms provide the reason for
the cancellation or alarm.
Diagnostic Cycle tapes indicate whether it passed or failed. If there was a failure,
it details the reason for the failure.
If the Operator ID, Physician ID, Device ID, Procedure ID, and/or Case ID options
are enabled, the cycle tape will also include the names or codes for these options.
If the Operator ID is disabled, a blank line will be printed on the optional paper
printout for manual entry.
Options
1. ID Options
The control (microcomputer-controlled Processor) of the Processor makes
available options for documentation of the following items associated with a
device being processed after a procedure:
• Operator ID
• Physician ID
• Device ID
• Procedure ID
• Case ID
Any or all of these can be enabled through Supervisor Mode (see Section 9). If ID
Options are enabled, the name or code will be documented on the electronic cycle
tape/printout. The method for entering the information is described in Section 7.
Operator Manual
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
Component Identification