Patient Leakage Current- Single Fault Condition (S.f.c) Mains On Applied Part - EDAN M3 Service Manual

Vital signs monitor
Hide thumbs Also See for M3:
Table of Contents

Advertisement

M3 Vital Signs Monitor Service Manual
(IEC/EN60601-1, UL60601-1)
Single fault condition: 500µA (BF applied part), 50µA (CF applied part)
(IEC/EN60601-1, UL60601-1)
Leakage Current
Earth Leakage Current
Enclosure Leakage Current
Patient Leakage Current
Patient Leakage Current (Mains on
Applied Parts)
Patient Auxiliary Current
3.3.5 Patient Leakage Current- Single Fault Condition (S.F.C)
Mains on Applied Part
NOTE:
The following test is based on test with the Fluke 601 pro series safety analyzer. This
device allows applying a 110% mains voltage between the applied part and the device PE.
When testing with other device, you may need to apply the 110% mains voltage manually.
Applied
Normal
Part
Condition
<0.5 mA
<0.1 mA
AC: <0.01 mA
CF
DC: <0.01 mA
AC: <0.1 mA
BF
DC: <0.01 mA
CF
BF
AC: <0.01 mA
CF
DC: <0.01 mA
AC: <0.1 mA
BF
DC: <0.01 mA
- 13 -
Test and Maintenance
Single Fault
Condition
<1 mA
<0.5 mA
AC: <0.05 mA
DC: <0.05 mA
AC: <0.5 mA
DC: <0.05 mA
<0.05 mA
<5 mA
AC: <0.05 mA
DC: <0.05 mA
AC: <0.5 mA
DC: <0.05 mA