Protective Earth Resistance; Enclosure Leakage Current - EDAN M3 Service Manual

Vital signs monitor
Hide thumbs Also See for M3:
Table of Contents

Advertisement

M3 Vital Signs Monitor Service Manual
Test and Maintenance

3.3.2 Protective Earth Resistance

NOTE:
The circuit diagram is based on the Fluke 601Pro series safety analyzer.
This measures impendence of Protective Earth (PE) terminal to accessible metal part of Device
under test (DUT) which is protectively earthed. A current of 25A is passed for 5s to 10s through
the protective terminal and each accessible metal part which is protectively earthed.
Allowable value: without mains cable, maximum impendence: 100 mOhms
(IEC 60601-1 and UL60601-1)

3.3.3 Enclosure Leakage Current

NOTE:
The circuit diagram is based on the Fluke 601Pro series safety Analyzer.
- 11 -