Note
The revision letter is located on the National Instruments label located on the side of
each SCXI-1120/D module. The revision letter is found in the part number and is the letter
found in that part number.
© National Instruments Corporation
The jumpers are used as follows:
Fixed jumpers
On Revision A and B SCXI-1120 modules, jumper W42 is unused
and should not be connected.
Jumper W45 is reserved and should not be reconfigured.
On Revision A and B SCXI-1120 modules, jumper W44 carries
the SLOT0SEL* signal from the rear signal connector, after
buffering, to the SCXIbus INTR* line and should be left in the
SCXI-1120 factory-default position (position 1). On all
SCXI-1120D modules and on Revision C and later SCXI-1120
modules, jumper W44 does not exist.
User-configurable jumpers
Jumper W43 carries the SCXIbus MISO line, after buffering, to
the SERDATOUT signal on the rear signal connector.
On all SCXI-1120D and on Revision C and later SCXI-1120
modules, jumper 42 connects a pullup resistor to the
SERDATOUT signal on the rear signal connector.
Jumper W46 configures the guard, the analog output ground, and
enables the NRSE mode.
Jumpers W1 through W8 configure the first-stage gain of channels
0 through 7, respectively.
Jumpers W9 through W16 configure the second-stage gain of
channels 0 through 7, respectively.
Jumpers W17 through W24 configure the first-stage filtering of
input channels 0 through 7, respectively.
Jumpers W25 through W40 configure the second-stage filtering of
input channels 0 through 7, respectively.
Jumper W41 directly connects the temperature sensor to the rear
signal connector. To do so, set jumper W46 in the AB-R2 or
AB-R0 position first.
Chapter 2
2-3
Configuration and Installation
SCXI-1120/D User Manual