Emptying the water trap
Remove the sample line from the water trap.
1.
2.
Press the grips of the water trap and pull outwards.
Insert an empty syringe
3.
20 mL without canula into the blue connector.
4.
Extract water, remove syringe and dispose of full syringe as infectious hospital waste.
Push the water trap into the holder or adapter again until it audibly clicks into place and
5.
connect the sample line.
WARNING
The liquid in the water trap may be contaminated and must be handled and disposed of with
care. Dispose of the liquid in an adequate way and in compliance with local regulations.
WARNING
If the sample line remains connected while the water trap is being removed, contaminated
liquid may escape from the water trap. Disconnect the sample line from the water trap before
removing the water trap from the medical device.
CAUTION
!
When the water trap is full, gas measurement can fail. Check the filling level of the water trap
regularly and empty it as necessary.
Dräger. Technology for Life
(C)
with a filling volume of at least
C
-20ml-
-20ml-
Risk of infection
Risk of infection
Risk due to gas measurement failure
®
Please note: This Quick Reference Guide is not a replacement or substitute for the
instructions for use. Any use of the medical device requires full understanding and
strict observation of the instruction for use