Installation - J.P. SELECTA DE-1626 Instruction Manual

Alcohol distillation unit
Table of Contents

Advertisement

Available languages

Available languages

INSTRUCTION MANUAL CODE 80191 REV D JAN/2020
J
.P. SELECTA s.a.u.

4 Installation

Once the unit is unpacked, visually check the packing list (see 1.1) and that there
are no parts damaged during transportation. Save the packing for some days for
later sending.
4.1 Suitable location
Place the unit over a flat level and stable surface, suitable for the unit weight.
No longer that 1.5m from a water tap, waste output and power supply plug.
To work comfortably, let a free space of 50cm on each side and not least than 20cm
on the rear part.
Prepare to refill the water reservoir from the top of the unit.
4.2 Connection to power supply
Locate a power supply socket (rated to 16A) near the unit.
Use the power cord supplied or another of similar specification.
For safety reason, the socket must have ground terminal, since the unit must work
grounded.
Before plugging on the socket, check that its voltage match the voltage marked on
the unit plate.
4.3 Connection to tap water and waste
Connect the water entry hose to a water tap (1). Connect the water output hose
from (2) to waste.
The hoses «remaining steam» (3) and «Steam generator empty» (4) must go to waste.
Do not forget the rubber white nuts on the hose's connections.
Press plastic connections by hand. Do not use any tool.
The tap water must have, at least, 1 bar pressure.
Keep the tap opened. The DE-1626 will open / close the cooling water automatically.
Note: It is critical in order that the unit works properly, the hose «Remaining steam»
(3) MUST NOT BE IMMERSED under water, or blended or blocked by any reason.
4.4 Water for steam generator
The unit includes a 2 litres water reservoir, at the top, to store the water for the
steam generator.
As the system used to control the level of water on the steam generator works by
conductivity, water of at least 20 microSiemens must be used. Water , also, must have
a low content of calcium to increase the life of the steam generator.
Water with this conductivity could be obtained from a demineralised unit. Or from
distilled water and add 5 mg/l Sodium Chloride.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(It can be modified without notice)
Tel 34 937 700 877 Fax 34 937 702 362
Page: 25

Advertisement

Table of Contents
loading

This manual is also suitable for:

4001626

Table of Contents