C7557-01 DLL INSTRUCTION MANUAL
5-5

AMP GAIN CONTROL

Controls the gain for the sensor output amp in the MCD controller. The return values are 1
with normal completion, and other than 1 with an error. The amp gain status may be
obtained as the return value for the function.
 Function name
WORD SetAmpGain(ID, AmpGain);
WORD GetAmpGain(ID);
 Format
WORD SetAmpGain(WORD, WORD);
WORD GetAmpGain(WORD);
 Arguments
ID
AmpGain
 Return code
For amp gain control
Same as for CheckSensor function
Status acquisition
10
USB ID number
Specify MCD controller USB ID number
Set amp gain (Low/Middle/High)
x1/2 : 0
x1 : 1
x10 : 4
x20 : 5
0 : x1/2
1 : x1
4 : x10
5 : x20
HAMAMATSU PHOTONICS K.K.
x2 : 2
x5 : 3
x50 : 6
x100 : 7
2 : x2
3 : x5
6 : x50
7 : x100