SETUP MENU (2/7)
10. IR filter
Turn Infrared filter off when using Infrared ring light.
To use regular ring lights turn the Infrared filter on.
11. White balance calibration at LED light change
Switch on if you want an automatic white balance calibration to be
performed when LED light is changed. Deactivated by default.
Switch off if you do not want an automatic white balance calibration to
be performed when LED light is changed.
12. Video format
Choose between various available video formats:
1080p60 / 1080p50 / 1080p30 / 1080p25. This choice is not available
in MOVE/FRONT, since this system runs 1080p60 per default.
13. Computer view
TREND/PRESTIGE only: Choose between Camera only (default)
and View all. If View all is selected, menus and applications will be
overlayed on the live video of the USB 3.0 output. This feature allows
you to use a computer to display the microscope live image and
screen share the output during video conferences.
MOVE/FRONT only: Choose between Stopped (default) and Streaming
to get an the live view from the camera displayed in a browser. To start
Streaming, please connect ethernet and access remoteview.tagarno.
com in a browser. Locate the session ID and enter the ID in the window
on the microscope. Please go to tagarno.com/remoteview-help for
further help and details.
14. Image format
Choose between different image formats when saving an image.
Available formats are TIFF, PNG, JPG or BMP.
NB: The following EXIF data are included when saving images in .jpg
format: Brand and model of microscope, creation data, keyword data
from apps, serial number and field of view. Any annotations added in
the Measurement application are also included here.
15. Image prefix
Replace the default image prefix (img_) preceding the date and time
stamp in filenames of captured images. This feature requires an
attached keyboard.
16. Assign image name
When this setting is active, the microscope will prompt for an image
name before storing a captured image.
NB! When using an attached scanner to name captured images on the
TAGARNO FHD microscopes we advise that the scanner is connected
directly to one of the two USB ports on the microscope.
It is not recommended to connect the scanner to an attached USB Hub
unless it has been thoroughly tested. The communication through some
hubs can unfortunately introduce data loss resulting in one or more
random characters missing in the scanned image name.
17. Image preview
This feature allows users to quickly inspect and capture images of new
samples. When an image is captured the default behavior is to preview
the image at the center of the display for 5 seconds. Image preview
can now be disabled, which will display a more discrete text box for 3
seconds.