Summary of Contents for LW MEASUREMENTS Tree LWC-P Series
Page 1
Professional Weighing Equipment Wireless Wheelchair Scale Operation Manual LWC-P Series Use factory adapter only NOTE: THIS PRODUCT HAS BEEN CALIBRATED IN THE FACTORY. PLEASE REFER TO THE MANUAL BEFORE RE-CALIBRATING PRODUCT! Rev 1.1, 18.09.2021...
Table of Contents Section Page Declaration of Conformity Customer Service Introduction Safety Weight Scale Functions Features Application & Conformity RS232 Data Getting Started Delivery Assembly & Installation Working Mode Cautions Application Menu Program Option 14-15 Calibration 16-17 Maintenance & Service Transportation &...
Imperial version Metric version LWC-M 400 LWC-M 800 Mark applied EU Directive Standards 2004/108/EC EN 61326-1: 2006 Signature: Boon Lim, R & D Manager Date: 28. 08. 2019 LW Measurements LLC, 620 Carlson Court, Rohnert Park, CA 94928 Rev 1.1, 18.09.2021...
Safety Representations and symbols Important instructions, which involve safety, are highlighted with the appropriate mark: Safety recommendations When using the weighing equipment in surroundings with increased safety requirements, the corresponding regulations must be observed. The weighing scale may only be used with the power adapter supplied. Before connecting the power adapter to the scale, the user must ensure that the operating voltage stated on the power adapter is compliant with the mains voltage.
Weight Scale Construction & Functions The weighing scale consists of the following parts; • The Scale body (1) • The Wireless indicator (2) • The adapter (3) • Operation manual. Figure 2.1 Your weighing machine Rev 1.1, 18.09.2021...
Functions The Large Wheelchair Scale is a high-quality electronic precision weighing machine with the following specifications Metric Specifications Model number Capacity Graduation Platter Size LWC-P 400 400kg 0.1kg 14.8 x 95 cm Shipping weight 17.5 kg Package 102 x 26 x 26 cm (Standard carton) 5 –...
Application & Conformity The weighing scale may only be used for the weighing of solid-materials and of liquids filled into secure containers. The maximum capacity load of the weighing scale must never be exceeded, otherwise the weighing scale may be damaged. In using the weighing scale in combination with other devices as well as with devices produced by other manufacturers, the appropriate regulations for the safe use of the relevant attachments and their application in accordance with instructions must be...
Getting Started The scale is packaged in an environmentally-friendly carton, which provides optimum protection for the scale during transportation. We suggest that you keep the original packaging in order to avoid damage if you are shipping or transporting the scale to a different location. It is also the best way to keep it in the best conditions if it will not be used for an extended period of time.
Assembly & Installation The weighing scale is delivered in a partly dismantled condition. Assemble the individual components in the following sequence: • Place your machine on a level, clean, and dry surface to obtain accurate readings. Connecting the AC Adapter The following Safety recommendations must be observed when connecting the scale to the mains: The Scale should only be connected to the mains voltage socket with the power adapter...
Working Mode Press the power switches of the two scales to turn on the power of the scales. Press [ON/OFF] key on the indicator to turn on the indicator. The bluetooth LEDS on the indicator and the scale body both change from flashing to bright, indicating that the indicator and the scale body are connected successfully.
Application Menu Display messages and key functions Display messages: Low battery indicator Negative value indicator Zero indicator Key Functions: • The ON/OFF key is to turn the Scale On or Off • The HOLD key is to store the weight value. •...
Program options Interface Settings − Press and hold the [HOLD] key, then press [ON/OFF] key to turn on the scale, the display will show A-ON or A-OFF. − Press [UNIT] key the display will show L-ON, L-Au or L-OFF. − Press [UNIT] key again to enter Parity mode, the display will show P-XXX. −...
Page 15
Setting power-on mode − Press and hold the [HOLD] key then press [ON/OFF] key to turn on the scale, the display will show A-ON or A-OFF. − Press [UNIT] key three times to the menu of setting the power-on display, and the display will show J-ON or J-OFF.
Calibration Using an External Calibration Weight Calibration is generally required when the weighing machine is initially installed or if the weighing machine is moved to a substantial distance from the original location. 30 minutes of warming up of the scale is always needed before initial calibration. A reminder when operating our larger capacity scales: The scales of the LWC-M series are calibrated at the factory.
12.Place the known test weight on the platform r, press the [UNIT] key after stable indicator displayed, it will show -----, and then will show CAL-r, the right scale calibration is complete. Press the [ON/OFF] key to turn the scale on to test if the weighing is accurate, if not, repeat above steps.
Our warranty does not cover misuse or neglect including but not limited to battery or water damage, overloading, and chewed or cut wires. If the product is found to have been misused or damaged by the customer, LW measurements is not responsible for the cost of return.
Need help?
Do you have a question about the Tree LWC-P Series and is the answer not in the manual?