5-12 S-METER ADJUSTMENT (continued)
1) Select an adjustment item using [2] / [8].
2) Set the SSG as specifi ed (frequency, deviation and output level).
3) Push the [BAND] to adjust (automatic) and store the adjustment value.
ADJUSTMENT
ADJUSTMENT
ITEM
S-METER S3 LEVEL
(WFM)
[1.01 MHz*]
[15.1 MHz*]
[40.1 MHz*]
[60.1 MHz*]
[220.1 MHz*]
[305.2 MHz*]
[435.1 MHz*]
[535.1 MHz*]
[800.1 MHz*]
; This output level of a standard signal generator (SSG) is indicated as SSG's open circuit.
*; This frequency may differ from the actual frequency displayed on the function display.
All manuals and user guides at all-guides.com
1 • Set the SSG as;
Frequency
Deviation
Level
2 • Set the SSG as;
Frequency
Level
3 • Set the SSG as;
Frequency
Level
4 • Set the SSG as;
Frequency
Level
5 • Set the SSG as;
[S3]
Frequency
Level
6 • Set the SSG as;
Frequency
Level
7 • Set the SSG as;
Frequency
Level
8 • Set the SSG as;
Frequency
Level
9 • Set the SSG as;
Frequency
Level
OPERATION
: (Displayed on the function display)
: 52.5 kHz
: 5 dBµ (–102 dBm)
: (Displayed on the function display)
: 3 dBµ (–104 dBm)
: (Displayed on the function display)
: 5 dBµ (–102 dBm)
: (Displayed on the function display)
: 5 dBµ (–102 dBm)
: (Displayed on the function display)
: 5 dBµ (–102 dBm)
: (Displayed on the function display)
: 5 dBµ (–102 dBm)
: (Displayed on the function display)
: 5 dBµ (–102 dBm)
: (Displayed on the function display)
: 5 dBµ (–102 dBm)
: (Displayed on the function display)
: 11 dBµ (–96 dBm)
5 - 13
VALUE
Push
[BAND].
(Automatic
adjustment)