Maintenance And Service; General About Service And Maintenance; Cleaning The Central Unit - Siemens LDS 6 Operating Instructions Manual

In situ laser gas analyzers
Table of Contents

Advertisement

Maintenance and service

6.1

General about service and maintenance

During normal use, the central unit LDS 6 requires no service. The optics housing with its optical
surfaces will need regular maintenance. Depending on application and purging method, the
interval of maintenance may vary from 1 to 12 months.
If the transmission in a channel drops below the level set by the user, the transmission alarm will
be activated. The optics housing in this channel needs to be serviced by cleaning the wedge
windows or realigning the optical path. For details on maintenance and service of the optics
housing delivered with your system, please refer to the corresponding manual.
Humid environment
Danger of electric shock.
● Avoid working on the device when it is energized.
● If working on an energized device is necessary, ensure that the environment is dry.
● Make sure when carrying out cleaning and maintenance work that no moisture penetrates
the inside of the device.
Dangerous voltage at open device
Danger of electric shock when the enclosure is opened or enclosure parts are removed.
● Before you open the enclosure or remove enclosure parts, de-energize the device.
● If maintenance measures in an energized state are necessary, observe the particular
precautionary measures. Have maintenance work carried out by qualified personnel.
6.2

Cleaning the central unit

NOTICE
Always make absolutely sure that no water gets into the unit during cleaning.
Failure to do so may result in a breakdown of the unit!
Use only a dry cloth without any cleaning agents. Since the central unit contains optical surfaces
great care should be taken when cleaning. Only outer surfaces may be cleaned.
LDS 6
Operating Instructions, 01/2022, A5E00295894-10
WARNING
CAUTION
6
63

Advertisement

Table of Contents
loading

Table of Contents