Quick Start
SMARTPAT ORP
Digital sensors for ORP measurement
Installation, assembly, start-up and maintenance may only be performed by appropriately trained personnel.
Check the nameplate for correct operating conditions.
For use in hazardous areas, special codes and regulations are applicable. Instruments must not be
connected to power supply before reading instructions described in the supplementary manual.
The responsibility as to the suitability, intended use and corrosion resistance of the used materials of this
device against the measured fluid rests solely with the operator.
SMARTPAT ORP sensors are factory calibrated. Nevertheless it is recommended to calibrate the sensor
before installing it into its final measuring location.
1. Connect the sensor to a device suitable for SMARTPAT ORP sensor calibration: OPTIBRIDGE / PC,
SMARTMAC, HART handheld, or your PLC.
2. Calibrate the sensor.
3. Install the sensor into its final measuring location.
1 Unpacking the sensor
Remove the protective cap from the sensor tip by gently twisting and pulling it .
• Lay the sensor on a soft ESD mat or soft paper tissue .
• Leave the protection cap  on the VP connector, as long as the sensor is not connected to the cable.
2 Electrical connection
Danger:
All work on the electrical connections may only be carried out with the power disconnected.
Take note of the voltage data on the nameplate.
Observe the national regulations for electrical installations.
04/2022 – 4009316501 - QS SMARTPAT ORP R01 en
For initial set up, we strongly recommend that you also refer to the relevant manual!
Connecting the cable to the sensor
For connection use only a VarioPin cable with shield wire, like the cable VP2-S.
 Ensure that both cable and sensor connector are absolutely dry
 Screw the cable connector on the sensor connector and tighten it by hand
Cable VP2-S
Transparent-black (inner coax shield)
Ub+
White
Ub-
Shield
S
Connection to junction box SJB 200 W(-Ex)
 SJB 200 W(-Ex) junction box, connection with or without internal resistor
 Sensor with VP2-S cable
 Display or other 4...20 mA loop powered device (e.g. data logger)
 Control system without 250 Ω resistor connected to internal resistor of SJB 200 W(-Ex)
 Control system with 250 Ω resistor connected to SJB 200 W(-Ex) without using internal resistor
SJB 200 W-(Ex) with internal resistor
S
Shield
A
Sensor +
B
Sensor -
C
Display +
D
Display -
E
Loop - w/ Display
G
Loop + (250Ω)
SJB 200 W-(Ex) without internal resistor
S
Shield
A
Sensor +
B
Sensor -
F
Loop - w/o Display
H
Loop +
1

Advertisement

loading
Need help?

Need help?

Do you have a question about the SMARTPAT ORP and is the answer not in the manual?

Questions and answers

Summary of Contents for KROHNE SMARTPAT ORP

  • Page 1  Screw the cable connector on the sensor connector and tighten it by hand 1. Connect the sensor to a device suitable for SMARTPAT ORP sensor calibration: OPTIBRIDGE / PC, SMARTMAC, HART handheld, or your PLC.
  • Page 2 Loop 1+ (HART resistor on board) Loop 1+ (external HART ® resistor required) Loop 1- Loop 1- Relay 1+ Relay 1+ Relay 1- Relay 1- Relay 2+ Relay 2+ Relay 2- Relay 2- 04/2022 – 4009316501 - QS SMARTPAT ORP R01 en...
  • Page 3 Save values? / select yes / press Enter / select no to deactivate manual hold / press Enter to leave calibration menu After unsuccessful message Warning!!! Offset out of range! appears behind the value New offset calibration 04/2022 – 4009316501 - QS SMARTPAT ORP R01 en...
  • Page 4 To avoid electrostatic charge, the metal enclosure shall be connected to the local equipotential bonding system. The digital Redox-Sensor, type SMARTPAT ORP 8150 must not be installed and operated in areas where excessive charge generating processes can occur (e.g. due to friction or oncoming flow with particles).