All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtechnik GmbH. Subject to change without notice. Copyright 2016 by KROHNE Messtechnik GmbH - Ludwig-Krohne-Str. 5 - 47058 Duisburg (Germany) www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
5.1 Calibration ........................23 5.1.1 Calibration with PACTware ....................23 ® 5.1.2 Calibration with HART Handheld 475 FIELD COMMUNICATOR ........25 5.2 Troubleshooting......................28 5.3 Status messages and diagnostic information..............29 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
Page 4
8.7 Overview Basic-DD menu tree (positions in menu tree) ..........43 8.8 Basic-DD menu tree (details for settings) ..............44 9 Appendix 9.1 pH as a function of mV....................47 9.2 pH temperature dependency..................48 10 Notes www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
In the table below, "x" is a placeholder for possible multi-digit alphanumeric combinations, depending on the available version. Release date SW/HW version Changes and Documentation compatibility 2015-12-01 SW 1.0.x MA SMARTPAT pH 8530 R03 HW 1.0.x 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
The manufacturer is not liable for any damage resulting from improper use or use for other than the intended purpose. The intended use of the sensor SMARTPAT PH 8530 is the measurement of pH value in liquids. 1.3 Certifications CE marking The device fulfils the statutory requirements of the following EC directives: •...
The manufacturer reserves the right to alter the content of its documents, including this disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable in any way for possible consequences of such changes. 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of icons as shown below. www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
In general, devices from the manufacturer may only be installed, commissioned, operated and maintained by properly trained and authorized personnel. This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
• SJB 200 W-Ex – Junction box Consumables/Spare parts available • Various pH buffer solutions for sensor calibration • Various cleaning solutions INFORMATION! For further information contact your local sales office. www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
Look at the device nameplate to ensure that the device is delivered according to your order. The sensor type is specified on the label of the sensor package and on the sensor itself. www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
Do a check of the packing list to make sure that you have all the elements given in the order. INFORMATION! Look at the device nameplate to ensure that the device is delivered according to your order. 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
A sensor specific DTM software for usage with PACTware FDT is available. The DTM software is free of charge and available from CD (scope of delivery) or can be downloaded from the KROHNE website (Downloadcenter). www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
1 Connect the sensor to the junction box or directly to the process control system. 2 Calibrate the sensor. 3 Install the sensor into its final measuring location. The required steps are explained in the following sections. 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
Conductivity Resistance Temperature Loop current The specification NaN (Not a Number) disappears after a few seconds once the initialisation of the sensor is completed. Afterwards, the measured values appear. www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
Connecting the cable to the sensor • Ensure that both cable and sensor connector are absolutely dry 1. • Screw the cable connector 2 on the sensor connector and tighten it by hand. 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
Page 19
INFORMATION! ® The SJB 200 W offers the opportunity to access the sensor via HART hand held. For further information refer to the manual of the SJB 200 W. 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
Basically any installation position is possible. However, ensure that sufficient medium flows through and around the electrodes (the conductive electrodes must always be completely surrounded by the medium). Structural measures must be taken to prevent flow interruption or gas bubbles. www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
This would interrupt the electrical contact between the inner buffer solution and the glas surface. Figure 4-6: Installation requirements 1 Measuring medium 2 Maximum deviation of 75° from vertical position 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
• Ensure that the raised face of the receiving part has a smooth surface • Screw the device into the thread by hand. Hand tight connection is sufficient. For this device the max. torque that can be used is 7 Nm. www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
The value 0 for the pH buffer solution appears. Enter the value of the pH buffer solution manually. • Press Next Next to proceed with the calibration procedure. Next Next • Clean the sensor with demineralised water 2. Figure 5-1: Calibration procedure 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
Page 24
Warning! Slope out of range! appears together with the message No saving No saving Warning! Slope out of range! Warning! Slope out of range! No saving No saving possible possible possible possible, no data can be saved. www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
0 pH and Buffer not recognised Buffer not recognised Buffer not recognised Buffer not recognised appear. • Press Ok Ok to proceed. Enter the value of the pH buffer solution manually. 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
Page 26
Ok to proceed. The message save values? save values? appears. save values? save values? • Select yes yes to save the values. • Press Enter Enter Enter to proceed. Enter www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
Page 27
Moisture on the sensor connector must be avoided! Moisture may cause a short-circuit and a malfuntion of the sensor! If moisture has entered the connector dry it with air (e.g. hot air gun). 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
Electrodes are exposed to air (not fully immersed). No temperature Temperature sensor is Exchange pH sensor. measurement faulty. (only applicable via ® HART DD or PACTware FDT/DTM). www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
Recalibrate the sensor. For more information refer Calibration with PACTwareTM on page 23. Slope > -50 mV/pH Zero point < -99 mV Zero point > +99 mV Maintenance interval expired 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
The zero point should lie between -58...+58 mV at pH 7. An optimal value is 0 mV at pH 7. The zero point of the sensor is displayed after each calibration procedure. The following figure shows the aging effects: www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
The manufacturer offers a range of services to support the customer after expiration of the warranty. These include repair, maintenance, technical support and training. INFORMATION! For more precise information, please contact your local sales office. 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
• such dangerous substances, to enclose a certificate with the device confirming that is safe to handle and stating the • product used. www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
The user must dispose of the WEEE to a designated collection point for the recycling of WEEE or send them back to our local organisation or authorised representative. 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
(KCl). The KCl solution itself is in electrical contact with the measuring medium by means of a diaphragm. The diaphragm prevents the measuring medium from penetrating into the reference system but still allows electrical contact with the measuring medium. www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
Page 35
To compensate for the temperature dependency of the pH measurement, the temperature of the medium can be measured and automatically compensated. Figure 7-3: Temperature dependency of the Nernst Factor (theoretical slope of a pH sensor) 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
Sensor shaft Glass Membrane glass A glass Inner buffer pH 7.0 Reference KCl gel Diaphragm 3 x ceramic O-Ring EPDM (FDA) Sensor head Nickel plated brass body with VP2 connector www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
Page 37
Acc. to EN 61326, NAMUR NE 21 compatibility: EMC Directive 2004/108/EC (valid until 2016/04/19) or EMC Directive 2014/30/EU (valid from 2016/04/20) INFORMATION! For further information contact your local sales office. 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
In the burst mode a slave device transfers cyclic pre-defined response telegrams, to get a higher rate of data transfer. ® There are two ways of using the HART communication: • as Point-to-Point connection and • as Multidrop connection, with 2-wire connection. www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
4 SMARTMAC 200 W 5 SMARTPAT Sensor ® 6 Secondary master with HART 7 Power supply for devices (slaves) with passive current output 8 Load ≥ 250 Ω (Ohm) 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
DDs on the Field Communicator. The Field Communicator must be equipped with a system card with "Easy Upgrade Option". For details consult the Field Communicator User’s Manual. www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
1 Process input 2 I/O ® 3 I/O HART 4 Device 5 Service 5 Service 1 Service calibration 5 Service 5 Service 2 Service parameter 3 Passwords 4 Sensor lock 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
5 Autoclaving counter Rise autoclaving counter (yes/no) 6 Input user password Set password 3 Logbooks 1 Calibration logbook Logbook for the last 20 calibration cycles 2 Error logbook Display errors www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
Page 45
Display PV - pH with min. and max. limits Display SV - electrode voltage in mV Display TV - temperature in °C / °F with min. and max. limits Display QV - resistance 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...
Page 46
NO to lock the sensor. If you select YES YES the setting is not reversible anymore and the sensor is ® invalid for use. The HART communication is not possible anymore. www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
At pH 7 and 25°C / 77°F the temperature error is zero. If temperature or pH changes the temperature error is calculated using the following formula: 0.03 pH difference / pH or 0.03 pH difference / K www.krohne.com 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en...
Page 51
NOTES SMARTPAT PH 8530 03/2016 - 4002544803 - MA SMARTPAT PH 8530 R03 en www.krohne.com...