How to Measure

How to Measure
1. Apply couplant to the surface of your part.
2. After power-up, place the probe FLAT on the
surface and press down.
3. Lift the probe when you hear a
double BEEP or see the green indi-
cator light blink. Measurement result
is displayed on the LCD.
A second reading may be taken at the same spot by
continuing to hold the probe down on the surface.
4. When completely finished, wipe the probe clean
of couplant then return both Gage and probe to the
protective case. There is no need to disconnect the
probe from the Gage during storage.
Zero Menu Option
Zero Menu Option
Zero
The probe must be periodically zeroed using the
ZERO menu option to compensate for both extreme
temperature and probe wear effects. Before using,
allow the probe to reach ambient temperature. Wipe
the probe clean of couplant.
If measurements will be made in extreme hot or cold
temperatures, it is recommended to ZERO the probe
in the working environment.
,
If measurements will be made on rough substrates
it is recommended to periodically ZERO the probe to
compensate for wear.