Monitor Relay Programming - Varian Turbo-V 2000HT Instruction Manual

Hide thumbs Also See for Turbo-V 2000HT:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 100
Press the PUMP CURRENT push-button to enter
the value and, if the serial option is installed, the
display shows:
R S 4 8 5
B A U D
R A T E
where: XXXX = means 800, 1200, 2400, 4800, 9600
baud
rate
for
the
communication. The controller is factory-set for 9600
baud rate operation.
Select the desired value by pressing the CYCLE
NUMBER.
R S 4 8 5
B A U D
R A T E
X X X X
host
computer
or
printer
9 6 0 0

Monitor Relay Programming

The display shows:
S P E E D
S E L E C T
where: XXKRPM = is the switch point of relay
the preset turbopump speed, adjustable from 00 to 99
KRPM.
The speed threshold will condition the K1 and K2
operation (see the following cycle diagram] and it is
factory-set to 30 KRPM. Press the CYCLE NUMBER
push-button to select the first number.
S P E E D
S E L E C T
85
TECHNICAL INFORMATION
T H R E S H O L D
I O N : X X K R P M
T H R E S H O L D
I O N : X X K R P M
87-900-913-01 (N)