TECHNO-GAZ ONYX B 5.0 Technical Manual page 97

Table of Contents

Advertisement

Available languages

Available languages

ENGLISH
PHASE 2 : FIRST PREHEATING (HEATING 1)
When EV4 is closed, the preheating phase with the working of the heating element
(positioned around the chamber) starts.
EV2
EV3
EV4
EV5
EV6
EV7
OFF
OFF
OFF
ON
OFF
OFF
The produced heat generates steam production inside the chamber, with conseguent raise of
the pressure, up to the value of + 0.3 bar.