PHASE 7 : THIRD VACUUM AND WATER INLET (VACUUM 3)
The autoclave activates the vacuum pump (F) and the status of EV5 changes in ON. Thanks
to the change of EV3 in ON, the air is suctioned from the chamber of sterilization. During the
phase of reaching of vacuum, EV6 activates and deactivates for cooling and for cleaning the
surface of the membranes of the vacuum pump.
EV2
EV3
EV4
EV5
EV6
EV7
ON (3 times at intervals -0.2bar /
OFF
ON
OFF
ON
OFF
-0.4bar / -0.6bar)
Once the negative pressure o -0.86 bar is reached, EV3 closes (off), the vacuum pump stops
working and the valve EV4 opens (on) to enable the inlet of water into the chamber from the
clean water tank (E). The quantity of water is read by the flowmeter; when the correct
quantity of water is reached, EV4 closes (off).
EV2
EV3
EV4
EV5
EV6
EV7
OFF
OFF
ON
ON
OFF
OFF
WATER TABLE INLET
Quantity of water for phase : ~350 / 470 ml
Total quantity of water : ~660 / 1000ml