A711
(in case the autoclave needs more than 45 min for the last heating)
Verify the correct working of the heating element and in case replace it.
Check that the safety thermostat doesn't be in the open position (to reset the correct
working, press on the push button which is on the thermostat)
Check the pneumatic circuit to verify pressure leakages
Check and in case replace the pressure transducer (procedure P17)
The MAIN BOARD doesn't read correctly the information sent by the transducer; replace it
T1 IN THE STERILIZATION'S PHASE OUT OF THE MIN
A751
(T1 out of the lower limit for + 10 sec / cycle 121°C ->limit 121°C - cycle 134°C -> limit = 134°C)
Execute from the technical menu: CALIBRATION PT1000 TEMPERATURE PROBES, to
Replace the PT1000 (temperature probes)
The MAIN BOARD doesn't read correctly the information sent by the PT1000, replace it.
The MAIN BOARD doesn't manage correctly the heater, replace it.
Check the pneumatic circuit to verify pressure leakages
T2 IN THE STERILIZATION'S PHASE OUT OF THE MIN
A753
(T2 out of the lower limit for + 10 sec / cycle 121°C ->limit 121°C - cycle 134°C -> limit = 134°C)
Execute from the technical menu: CALIBRATION PT1000 TEMPERATURE PROBES, to
Replace the PT1000 (temperature probes)
The MAIN BOARD doesn't read correctly the information sent by the PT1000, replace it.
The MAIN BOARD doesn't manage correctly the heater, replace it.
Check the pneumatic circuit to verify pressure leakages
LAST PREHEATING
TECHNICAL INTERVENTION
TECHNICAL INTERVENTION
proceed with the re-calibration of the PT1000 probes
TECHNICAL INTERVENTION
proceed with the re-calibration of the PT1000 probes
47
LIMIT
LIMIT