PRECISION MEASURING TOOLS
LEFT HAND DIGITAL VERNIER CALIPERS
Thank you for purchasing Niigata Seiki Digital Calipers.
This is a precision measuring tool for making Inside, Outside, Step Height, and Depth measurements.
●For safe and proper use, please read this instruction manual before use and follow the procedures described.
Please keep manual where it is accessible to user for future reference.
●Keep this manual with the instrument if transferred or leased to a third party.
●For inquiries about this product, please contact dealer or Niigata Seiki.
■PART IDENTIFICATION AND FUNCTION
※Model D-150HL shown.
⑧Scale
⑥ Depth Bar
⑪ Battery Cover
⑦ Measuring surfaces
for depth measurements
①Measuring Surfaces for outside measurements ...
②Outside Jaws .............................................
③Measuring surfaces for step measurements ......
④Inside Jaws................................................
⑤Measuring surfaces for inside measurements ......
⑥Depth Bar ................................................
⑦Measuring surfaces for depth measurements ......
⑧Scale ......................................................
⑨Lock Screw .............................................
⑩LCD Display .............................................
⑪Battery Cover ..........................................
⑫ [ ZERO ] Button ..........................................
⑬ [ ON/OFF ] Button .......................................
SAFETY PRECAUTIONS
To prevent harm to yourself or others, and to prevent damage to
property, always follow the procedures marked by the following
symbols.
WARNING
Indicates risk of personal injury or property damage if not followed.
Read the manual and follow the directions.
・Use of product other than as described in this manual
may cause accident.
Use only for measuring.
・Use for any purpose other than measuring may
damage or wear the instrument. Improper use may
also cause accident.
Handle with Care.
・Do not drop or subject instrument to excessive shock.
Do not place under heavy objects. Improper handling
may cause damage or poor accuracy.
・Do not scratch instrument, for example by writing ID
number.
Avoid using in high electric fields, such as near
fluorescent lights or switching power supplies.
・Incorrect value may be displayed on LCD.
User Manual
Model No.
⑤ Measuring surfaces for
inside measurements
⑨Lock Screw
⑩LCD Display
④Inside Jaws
③Measuring surfaces
② Outside Jaws
⑫ [ ZERO ] Button
⑬ [ ON/OFF ] Button
① Measuring surfaces for
outside measurements
Object to be measured is positioned between these surfaces.
For measuring outside dimensions.
Measures height of steps.
For measuring inside dimensions.
Position surfaces inside object to be measured.
For measuring depth.
Depth is measured between these two surfaces.
Supports LCD Display Unit.
Tighten to Prevent the Slide and Display from moving.
Displays digital measurement.
Open to access battery.
When pressed, Display will read "0.00".
For turning power ON /OFF.
Please Observe
Denotes a prohibition -
Denotes a requirement -
You MUST NOT do
You MUST do
Use and store only in environment meeting the
following conditions:
●Dry location protected from rain, water, and oils.
●Location that is not excessively hot.
●Location protected from use by children and
unauthorized people.
・Use in places contrary to the above may cause
damage to the product resulting in accident or injury.
Jaw Tips Are Sharp̶Handle With Care.
・Careless handling may cause injury.
Do not disassemble or modify.
・Disassembly may damage product or cause accident.
For repair, please contact distributor .
Do not write on instrument with electric engraver.
・Do not engrave instrument body or jaw, such as an ID
number with engraver or electric pen.
PREPARATION FOR USE
D- *** HL
Before using the calipers, please follow the steps below:
●Wipe off rust proofing and dirt from measuring
surfaces and slide.
Contamination of surfaces may cause measurement
error.
●Allow the calipers and object to be measured
time to reach the same temperature.
A temperature difference between the object to be
measured and the caliper may cause measurement
error. Allow enough time for the temperatures to
equilibrate.
INSERTING・REPLACING BATTERY
① Slide cover in direction of arrow.
for step measurements
② Insert battery with (+) side outward.
③ Replace battery cover.
Notice
Use only SR44 type battery.
Battery
Cover
Remove
TROUBLESHOOTING
Q. Nothing is displayed on LCD.
A. The battery may be running low, please replace
battery with a new one.
Q. With power on, the number on the LCD does not
change as Display Unit is moved.
A. Please reset by removing battery for about 30sec.,
and then replacing.
Q. Nothing happens when buttons are pressed.
A. Make sure battery is properly inserted.
Q. The correct value is not displayed.
A. High humidity may cause dampness. Please dry off
the instrument and use only in dry location.
SERVICE
●If Instrument is not working properly, or if you have any
questions, please contact distributor.
●Please note, manufacturer is unable to respond to
inquires or provide service directly. Please contact
distributor .
HOW TO USE
■Power ON/OFF
①To turn on, press the [ ON/OFF ] Button.
②To turn off, press the [ ON/OFF ] Button.
※
Power will automatically turn OFF after about 20 minutes with no
activity. Press the [ ON/OFF ] Button or move the Slide to turn ON.
■Absolute Measurement Mode
①Loosen the lock screw, and turn on power.
②With jaws closed, press the [ ZERO ] Button.
③When "0.00mm" is displayed on LCD, Calipers is ready to
start measurement.
■Comparisons Measure Mode
①Loosen the lock screw, and turn on power.
②With jaws open to desired reference position ( any point
is OK, ) press [ ZERO ] Button.
③When "0.00mm" is displayed on LCD, Calipers is ready
to start measurement.
AFTER USE CARE・STORAGE
●After use, wipe off any dirt or oil from
measuring surfaces, sliding surfaces, and
exterior using a dry cloth and perform rust
prevention treatment.
For rust prevention, lightly apply anti-rust oil, or place in
rust preventive bag.
●Store with Jaws slightly open.
If Jaws are closed, expansion due to temperature changes
may cause excessive force which will cause inaccuracy.
●Store in a cool, dry, and dark location in the
provided case.
Keep out of direct sunlight and moisture, and please
keep secure from unauthorized personnel.
●Remove battery if Caliper will not be used for a
long period of time.
Batteries may leak if left in unused instrument for an
extended period.
PRODUCT SPECIFICATIONS
※
Measuring
Maximum
Model No.
Range
permissible error
D-150HL
0 -150
mm
±0.03
mm
D-200HL
0 -200
mm
±0.04
D-300HL
0 -300
mm
mm
※Not including quantization error ( ±1 count )
●Resolution : 0.01mm
●Display : 5 Digit LCD
●Auto Off : approx.20 min.
●Maximum response speed : Slide to Right .....1.5 m/s
Slide to Left .......0.75 m/s
※Left/Right as shown in diagram at top of manual.
●Operating Temperature Range :
0° C~40° C
●Storage Temperature Range : −10° C~ 60° C
●Power : ( 1x ) SR44 Silver oxide Battery
●Battery Life : approx. 1 years in normal use
●Accessory : SR44 Silver oxide Battery ( for test )
Niigata seiki Co., Ltd.
5-3-14, Tsukanome, Sanjo, Niigata, Japan, 955-0055
Tel. : +81-256-33-5522 Fax. : +81-256-33-5518
MAIL intl.sales@niigataseiki.co.jp
URL
http://www.niigataseiki.co.jp
Weight
150g
190g
250g