Daily/Weekly Testing

Manufacturer's testing MUST be conducted as outlined
below. We recommend the completion of our Washer
Disinfector Log Book (Part number: 119858) - refer to 8.2
Spares, for ordering details.
Daily testing
Clean the door seal and cabinet face using Eschmann
recommended wipes (Contact Eschmann).
Check spray arm nozzles for blockages.
Check spray arms rotate freely (This test should also be
conducted every cycle).
Clean the cabinet strainer and filters face using
Eschmann recommended wipes (Contact Eschmann).
Perform a visual check of load cleanliness (This test
should also be conducted every cycle).
Weekly testing
Check that the door locks and microswitch are operating
correctly.
Check the physical condition of the door seal.
Perform a test for cleaning efficacy and residual proteins:
1.
Place a wash clean indicator (8793193) into the
wash clean holder (8793228).
2.
Place the wash clean test device into the washer
disinfector as shown below. A normal load can be
processed at the same time as the test takes place,
but make sure that the load does not obstruct the
flow of water from the spray arms to the wash clean
test device.
3.
Run either the Eschmann Dental or W T D cycle.
4.
When the cycle has completed, remove the wash
clean test device.
5.
Check the wash clean indicator - it should be visibly
clean. Swab this indicator with a protein test pen
(3371) to check if any protein is present on the
indicator.
Note: Be careful not to touch the wash clean
indicator or protein test swab because the test
may fail.
6.
Review and store the cycle printouts (Paper or
digital) to confirm that all test cycles have passed the
test.
Position of wash clean holder on bottom carrier.
MUW216743-2
Wash clean holder (8793228)
Wash clean indicators (8793193)