Seca 769 LC Service Manual page 11

Colum scale with 4 load cell technology

Advertisement

Service Manual
Scales with LC electronic
Please refer to the operating instructions to ensure that the scales were operated in the
proper manner. In case of a malfunction, the following table will help you to locate the fault.
Description of the fault
Scales cannot be started
Scales show no weight or
an incorrect weight
(repeatedly)
Scale shows only the
segment test or the
readings are varying to
much.
Incomplete digits
Malfunction during the
weighing procedure
E XX.11
E XX.12
E XX.13
E XX.16
E XX.40
E XX.50
E XX.60
19.07.07 /Kr/Rei
Possible cause
Battery is not
installed or
discharged
Voltage supply is
defective
Bus connection
cable is defective
Keypad is defective
or not connected
Bus fault
Packaging /
transport safeguard
has not been fully
removed
Scales are
maladjusted
Sensor is damaged
transfer of force to the
sensor is interrupted
Shortcircuit between
sensor cable and
housing
Display is defective
Display:
Er:x:no
Sensor is defective or
supply lines are
damaged
• Check the initial load
Failure at
Identification
initial zero setting
Limit for zero
tracking is reached
Failure at
Identification
initial zero setting
No EEProm data
Flash memory
damaged
Bus connection
cable is defective
Check the batteries
Check the voltage at the appropriate connections of the electronics;
eliminate the interruption
Check the plug-in connections; check the cables for interruption
and replace them if necessary.
Check the cable connections; check the cables and the keypad for
interruption and replace them if necessary
Disconnect power supply unit and battery from the scale, wait for 60
seconds, then reconnect.
Check the scales
Recalibrate the scales
Replace the sensor and start an additional corner override
Check the base frame for damage or incorrect assembly
Search for damaged isolation of the cabel and repair it. ( If
appropriated use multimeter.)
Replace the electronics
The electronics have detected an error.
n X:
module number
n No: error number
Check connection cables, check sensor and replace as required (if
parts a replaced, a corner override is necessary)
If several load cell are used, check the allocation (colour codes)
Sensor defective, replace it possibly (a corner override is
necessary)
Load the scale at work with >50g/s
while the start process beware the scale for agitation
Recalibrate the scale
Replace the electronics
Check the plug-in connections; check the cables for interruption and
replace them if necessary, Disconnect power supply unit and
battery from the scale, wait for 60 seconds, then reconnect.
Description of faults
Remedy
30-34-00-744