Service manual
7.
Once the scale has identified the load in the middle of the weighing platform, "CAL L"
comes up on display and, some seconds later, the display starts flashing to request
that a load be placed on the scale. The correct weight must now be acknowledged by
briefly pressing any of the buttons (with the exception of the start button). Once a
stable weight value has been recognized, the LCD stops flashing and a calibration
value is calculated and saved. Shortly afterwards, the text "Sto" is displayed on the
LCD and the scale automatically switches off some seconds later.
8.
Then remove the weight from the scale, return the slide switch on the PCB to its
original position and re-assemble the display housing.
9.
Now set the scale up again and then switch it on. Afterwards, place a calibrating
weight on the scale and check whether the correct weight is displayed. If recalibration
was successful, secure the display housing again with a seal in accordance with
national regulations.
1.2.2 Aborting a recalibration
The recalibration function can be aborted by switching the scale off before the requested weight has
been acknowledged
12.02.08 /Kr , Rei
2 (2)
Calibration
30-34-00-743_E1