3-3-2. Checking the needle echo

(1) Dip the probe tip into the warm water so that the needle echo is displayed.
(2) Check the following points.
• The needle echo matches with the puncture guide line.
• The echo of the entire needle is displayed fully and clearly.

3-4. Performing cleaning and sterilization

(1) Before use, clean, disinfect and sterilize the puncture guide tube.
See manual of puncture guide tube MP-2452.
(2) Clean, disinfect and sterilize the probe to be used in accordance with its usage purpose.
Warning
Clean, disinfect and sterilize before using the equipment.
Perform proper cleaning, disinfection and sterilization after use.
Otherwise, there is a risk of infection. Note that the equipment is not sterilized at the factory.
Before using the equipment first, be sure to clean, disinfect and sterilize it.
Probe
Images on the monitor
Puncture guide tube
Water of 40℃
-22-
Tub
Needle echo
Puncture guide line