Point Of Use (Pre-Cleaning); Containment And Transportation - Hitachi UST-676P Instruction Manual

Transrectum puncture electronic convex probe
Table of Contents

Advertisement

5-3. Point of use (Pre-cleaning)

In the operating room after use of the probe
A). Probe
1) Remove any accessories from the probe like puncture guide tube and rubber boot.
2) Flush patient's blood or fluid by tap water directly after use until the surface looks visually clean.
3) Wipe the whole surface of the probe by gauze pad and remove superficial visible impurities until the
surface looks visually clean.
B). Accessories
1) Clean the accessory of all patient's blood or fluid with running tap water until the surface of the
accessory looks visually clean.
2) Wipe the whole surface of the accessory by gauze pad.

5-4. Containment and transportation

Putting the contaminated equipment into exclusive shock and damage proof container for transportation is
recommended. It is recommended that instruments are reprocessed as soon as possible and not later than 4
hours after usage.
Caution
Do not immerse the probe in any liquids beyond the range of IPX7 shown in section 2-2 "Specifications".
Use when liquid has gotten inside the connector can result in a risk of electric shock to the operator
or patient. If liquid gets inside the connector, immediately stop use and contact one of our offices and/
or distributor's offices listed on the back cover.
-37-
MN1-5029 Rev.22

Advertisement

Table of Contents
loading

Table of Contents