Periodic Electrical Testing

Electrical Safety Tests must be performed by qualified personnel authorized
by the institution to perform such testing. The Safety Tests must be
performed and documented at least once per year, or according to
institutional policy. These tests include but are not limited to:

• Leakage current

• Ground bond test

Leakage Current
Leakage current must be tested according to methods and pass/fail
criteria described in IEC 60601-1:2005. Leakage current must be
performed with the heater circuit in the full ON condition. To achieve
this condition, perform the test when the reservoir is at room
temperature. When the HOTLINE
temperature is rapidly rising, but still below 39°C, the heater circuit is
in a full ON condition.
Note: The HOTLINE
HOTLINE
Fluid Warming Set is required to correctly operate the
HOTLINE
Warmer and perform leakage current testing. Do not defeat
the sensing interlocks or try to operate the HOTLINE
HOTLINE
Fluid Warming Set in place.
Ground Bond Test
Ground bond test must be tested according to methods and pass/fail
criteria described in IEC 60601-1:2005.
®
HOTLINE
Blood and Fluid Warmer | Operator's Manual
S E C T I O N 9
Warmer is first turned on and the
Warmer is equipped with sensing interlocks. A
T e s t i n g
without a
27