Step 4 - Fill The Reservoir With Recirculating Solution; Step 5 - Perform The Electrical Safety Tests - Smiths Medical LEVEL 1 HOTLINE HL-90-40 Operator's Manual

Blood and fluid warmer
Table of Contents

Advertisement

a
1.4 LITERS
S E C T I O N 4

Step 4 - Fill the Reservoir With Recirculating Solution

• Do not fill the HOTLINE
Fluid Warming Set in place. Failure to remove the HOTLINE
Fluid Warming Set before the fill procedure may result in an air
lock in the HOTLINE
Recirculating Solution Protocols
Use one of the following solutions for the reservoir.
Recirculating Solution
Preparation
0.3% Hydrogen Peroxide
Mix 140 ml of 3% hydrogen
Solution
peroxide with 1,260 ml of
distilled water.
Distilled Water
Use distilled water.
35% Isopropyl Alcohol
Mix 700 ml of 70% isopropyl
Solution
alcohol with 700 ml of distilled
water.
Note: Use distilled water only, not tap water. Failure to do so may cause build-up of
mineral deposits in the recirculating solution path, which may impair heater
performance.
1 Prepare the recirculating solution.
2 Remove the fill-port plug ( a ).
3 Fill the reservoir with 1.4 liters of recirculating solution.
4 Replace the fill-port plug.

Step 5 - Perform the Electrical Safety Tests

Perform all applicable electrical safety tests as required per institutional
procedure. Refer to Section 9,
electrical safety testing.
®
HOTLINE
Blood and Fluid Warmer | Operator's Manual
A s s e m b l y I n s t r u c t i o n s
WARNINGS
WARNINGS
Warmer reservoir with a HOTLINE
Warmer.
Testing
, for more information about

Maintenance
Replace solution and
disinfect reservoir every
12 months.
Replace solution and
disinfect reservoir every
30 days.
Replace solution and
disinfect reservoir every
30 days.
11