Add Recirculating Solution Mode; Over Temperature Alarm Mode - Smiths Medical LEVEL 1 HOTLINE HL-90-40 Operator's Manual

Blood and fluid warmer
Table of Contents

Advertisement

To clear this mode, check that the Twin-Tube Connector on the
HOTLINE
Fluid Warming Set is firmly inserted in the socket.

Add Recirculating Solution Mode

The Add Recirculating Solution mode indicates that the solution level
in the reservoir is below its minimum level.
Mode characteristics
• The green Operating LED on the Display Panel turns off.
• The red Add Solution LED ( a ) on the Display Panel illuminates.
• The audible alarm sounds and repeats approximately every two
seconds.
• The recirculating solution stops circulating.
To clear this mode, add recirculating solution to the reservoir.

Over Temperature Alarm Mode

The Over Temperature Alarm mode indicates that the temperature of
the recirculating solution is at or above 41.2°C.
Mode characteristics
• The green Operating LED on the Display Panel turns off.
• The red Over Temperature LED ( b ) on the Display Panel
illuminates.
• The audible alarm sounds and repeats approximately every two
seconds.
• The recirculating solution stops circulating.
For instructions to clear this mode, see Section 8,
16
HOTLINE
Troubleshooting
®
Blood and Fluid Warmer | Operator's Manual