To verify the recirculating solution temperature, you will need the
following:
• TEMP CHECK (HLTA-40)
• HOTLINE
Warmer
• HOTLINE
Fluid Warming Set
To Verify the Recirculating Solution Temperature:
1 Plug the HOTLINE
2 Place the TEMP CHECK on the top right corner of the
Warmer and plug it into the socket on the right side of
HOTLINE
the HOTLINE
Warmer.
3 Plug the Twin-Tube Connector on the HOTLINE
Set into the socket on the right side of the TEMP CHECK.
4 Remove the black label from the auxiliary outlet on the back of the
HOTLINE
and plug in the TEMP CHECK power cord.
Note: The auxiliary outlet is for use only with Smiths Medical
accessories.
5 Turn ON the HOTLINE
temperature to stabilize.
6 If the TEMP CHECK display indicates a temperature between
39 ° C and 40 ° C, and the HOTLINE
TEMP CHECK display, recirculating solution verification is
complete. Refer to the TEMP CHECK Manual for OVERTEMP
ALARM verification.
7 If the TEMP CHECK display does not indicate a temperature
between 39 ° C and 40 ° C, refer to the TEMP CHECK Manual for
calibration instructions.
26
HOTLINE
Warmer into a power outlet.
Warmer. Allow 15 minutes for the
Warmer display equals the
®
Blood and Fluid Warmer | Operator's Manual
S E C T I O N 9
Fluid Warming
T e s t i n g