Required Differential Pressure - Bopp & Reuther DIMF 2.0 TVS Operating Instructions Manual

Density transducer
Table of Contents

Advertisement

1.3 Required differential pressure

Density transducers of the DIMF series measure independently of flow rate and also at flow rate zero.
Their application is therefore normally problem-free. Care must, however, be taken to ensure that the
operating flow rate in the transducer
- updates the sample fast enough
- equalizes the temperature in the tranducer
- avoids air or gas bubbles or
deposits in the oscillating tube
- does not cause cavitation in the oscillating tube
- does not cause wear through abrasives
Recommended operating flow rate through the density transducer:
Pressure loss diagram
Pressure loss
10
∆P [bar]
(H
O; 20°C)
2
1
0.1
0.01
0.001
0.1
0
DIMF2.0
0.1
We reserve all rights of ownership and exploitation in respect of these
documents, including industrial property rights. These documents may only
be used with our express consent in writing, and only to the extent permitted
in any such consent. The documents may not be duplicated or made
available to third parties. In case of violation of the aforementioned
provisions, we reserve all rights.
DIMF 2.0: approx. 2.0 l/m.
DIMF2.0
1
10
1.5
6
2 l/min
1
10
0.01
0.1
Flow rate [l/min]
100
50 l/min
Permissible flow rates
Recommended flow
l/min
100
m³/h
1
10
7