Evaluation Electronics Of Transmitter Type Tr - Bopp & Reuther DIMF 2.0 TVS Operating Instructions Manual

Density transducer
Table of Contents

Advertisement

1.2 Evaluation electronics of transmitter type TR

Functions
Display parameters
Programmable parameters
HART
protocol
Output signal
Power supply
Connection
Cable specification
Ambient temperature
Storage temperature
Safety classes
Degree of protection (housing)
Dimensions (housing)
Material (housing)
Weight
Calibration and configuration
We reserve all rights of ownership and exploitation in respect of these
documents, including industrial property rights. These documents may only
be used with our express consent in writing, and only to the extent permitted
in any such consent. The documents may not be duplicated or made
available to third parties. In case of violation of the aforementioned
provisions, we reserve all rights.
Excites the oscillating element of the density transducer to
its natural frequency.
Equipped with a two-line display and four keys for displaying
data and configuring the on-site transmitter.
HART
communication
When the process data are changed, the user can perform a
simple modification of the set parameters.
Density, concentration, operating temperature, etc.
Lower range and upper range value of the output signal
(smallest measuring span approx. 5 kg/m³)
Calibration constants, constants of liquid, reference
temperature, etc.
Operation using a PC or laptop equipped with the
SensorPort 2 configuration software together with the HART
interface or operation using a HART
2.3 HC-275 of Rosemount
4-20 mA, linearized and temperature-corrected,
can be assigned to any display parameter
(e.g. operating density, reference density, concentration, °Brix,
°Plato or other magnitudes derived from density)
24 V DC (min. 14 V DC / max. 30 V DC)
2-wire technique via screw terminals;
cable enters via cable gland with M 20 x 1.5
or ½" NPT thread for pipe installation (conduit system)
2 wires, twisted and shielded
-10°C to +58°C
- 40°C to +70°C on request
-40°C to +70°C
Exi (standard version):
EEx ia IIC T4
ZELM 99 ATEX 0008 X
Device group II
Category 1/2 G
Measuring tube designed for Zone 0
Exi (Tantalum version):
EEx ia IIC T4
ZELM 99 ATEX 0008 X
Device group II
Category 2 G
Exd:
see individual approval
94/9/EG, FM, CSA in preparation
IP 65
ø100 (D) x 155 (L) x 120 (H) mm
Cast aluminum
1.2 kg
According to customer data at
Bopp & Reuther Messtechnik GmbH factory
hand-held terminal
6

Advertisement

Table of Contents