Electromagnetic Compatibility (EMC) Information
Interference immunity test
Electrostatic discharge in accordance with
IEC 61000-4-2
Electrical fast transients / bursts in accor-
dance with
IEC 61000-4-4
Surges
Line-to-line in accordance with
IEC 61000-4-5
Surges
Line-to-earth in accordance with
IEC 61000-4-5
Voltage dips in accordance with
IEC 61000-4-11
Voltage interruptions in accordance with
IEC 61000-4-11
Power frequency magnetic field immunity in
accordance with
IEC 61000-4-8
1568913, Edition 2019-06, Version 0
14.1.2 Electromagnetic immunity
The Navigator Lift
™
180 is intended for use in the ELECTROMAGNETIC environment
specified below. The customer or the user of the Navigator Lift
is used in such an environment.
Test level in accordance with IEC 60601
±8kV contact
±2kV, ±4kV, ±8kV, ±15kV air
±2kV
100kHz repetition rate
±0.5kV, ±1kV
±0.5kV, ±1kV, ±2kV
0% U
; 1/2 period
T
at 0, 45, 90, 135, 180, 225, 270 and 315 degrees
0% U
; 1 period
T
and
70% U
; 25/30 periods
T
Single phase: at 0 degrees
0% U
; 250/300 periods
T
30A/m
50Hz or 60Hz
Please note:
U
is the a.c. mains voltage prior to application of the test level.
T
50
™
180 should ensure that it
Test result
passed
passed
passed
passed
passed
passed
passed
passed