O & M Manual Model Q46 Ethernet-IP Communications Manual Home Office European Office Analytical Technology, Inc. ATI (UK) Limited 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park Collegeville, PA 19426 Delph New Road, Delph Phone: 800-959-0299 Saddleworth OL3 5DE...
Figure 7 – Successful Detection of MAC ID and IP Address ......... 12 Figure 8 – PING and ARP Command Tests for IP Address Setting......13 Figure 9 - Q46 Ethernet/IP Register Data Format ............15 Figure 10 – Base Status Register Bitfield Flags ............15...
Any communication option can be plugged into any Q46 for immediate conversion to a new protocol. Because this gateway device is part of the Q46 interface, the ID and device type show up on the interface as the gateway device. However, the Q46 variation...
Ethernet/IP Card Installation Instructions Use the following instructions to install the Ethernet/IP Communications Option into an existing Q46 Analyzer. Monitors ordered with the Ethernet option will be supplied with the board already installed. 1. Disconnect power to the Q46 prior to installation of the board.
Industry Association. It is readily available in lengths up to 100 ft. (30 m) with plugs on each end. To install an Ethernet cable in the Q46, pass the unterminated cable through the cable gland nearest the location of the RJ45 connector on the Ethernet/IP option board.
IP address of a Q46D. This free Molex PC tool can be used to test explicit messages on the Q46 interface. To run this on a PC for check-out, the user must first be sure to set the IP address of the PC Ethernet port to the 192.168.0 subnet range, like 192.168.0.20 (or, manually run a BOOTP...
Page 8
Class = 4 Instance = 101 Attribute = 3 The raw response shown above from the Q46 is shown below - 8E, 00, 00, 00, 00, 00, 00, 00, 07, 01, 00, 00, BC, 09, 00, 00, 3E, 01, 00, 00,...
Configuring fixed IP with BOOTP The Q46 acquires a dynamic network IP address from the network client/master through the BOOTP service, which is enabled by default for ODVA EIP-CIP conformance. However, a fixed IP can be set using BOOTP on a PC.
PC to a subnet that can reach the existing Q46. Turn off Q46. Connect Q46 to laptop with direct cable. Set laptop NIC port on a locked IP and subnet to stop DHCP service. This is critical. In IPV4 properties on your PC’s Ethernet connection port, select "Use the...
PC will now be able to access any network device on 192.168.0.X, as the subnet mask 255.255.255 locks in the first three variables. The default value for the Q46 is 192.168.0.254, so it can now be reached by the PC. Find the Rockwell BOOTP PC application and launch it.
Figure 6 – Setting IP/mask Assignment in BOOTP Server Now, plug the cable from Q46 into the PC Ethernet port, turn on power to the Q46, and wait for BOOTP to detect the request from the Q46 for an IP address.
Then close out all BOOTP server windows. Now, to check that the IP address has been properly saved, apply power to Q46 again and make sure it is connected to PC Ethernet port. Do not launch BOOTP server this time, as you don’t need it.
Page 14
Q46 Ethernet/IP Communications Manual as this is the fall-back IP that has been stored in the Q46. On power up, the Q46 will still try and find a BOOTP server for 30 seconds, and then default to this stored value as the fallback. If you do not see your new connection, two troubleshooting notes –...
The data organization shown in section 1.6 corresponds to the same mapping as all the other communication interfaces on the Q46. All Q46 instruments provide 20 bytes of input data to the master, and the meaning of the bytes changes slightly depending on which Q46 instrument is used.
Q46 Ethernet/IP Communications Manual Main Data Aux1 Data Aux2 Data Instrument Main Aux1 Aux2 Format Format Format Q46H/62/63 14920= 700= 475= /79PR Chlorine 1.4920 ppm 7.00pH 47.5% ID=41 Q46S/66 14920= 475= Sulfite None None ID=81 1.4920 ppm 47.5% Q46S/81 14920=...
Page 17
PRODUCT WARRANTY Analytical Technology, Inc. (Manufacturer) warrants to the Customer that if any part(s) of the Manufacturer's equipment proves to be defective in materials or workmanship within the earlier of 18 months of the date of shipment or 12 months of the date of start- up, such defective parts will be repaired or replaced free of charge.