2.0 Introduction to the Unit

2.1 Overview

The CSM (Copper Silver Monitor) has been specially designed to
measure and report the concentration of copper and silver dosing
in water samples over specified ranges from an ionization system.
The monitor uses chemistry to test for copper and silver at the
required concentrations.

2.2 Specifications

NOTICE
Specifications listed are based on installation/operation at sea level.
Measurement Range
50 ppb to 250 ppb silver
200 ppb to 2000 ppb copper
Repeatability
±10% RSD
Accuracy
±20%
Time Between Measurements User selectable up to 23 hours, 59 minutes
Path Length
10mm

Display

3.5" Color graphic, touch screen
Analog Output
Powered 4-20mA, 1000 Ω drive, isolated
Maximum Water Pressure
124 kPa (18 psi) (see Plumbing Connections section)
Flow Rate
185 ml/min. to 225 ml/min. (.044 to .059 Gal/min)
Operating Temperature
1°C – 50°C (34°F – 122°F)*
Max Sample Water Intake
225 ml/minute
and Drainage Flow
Wetted Materials
PBT, Polypropylene, 316 Stainless Steel, Viton®, Noryl
Reslyn, Borosilicate Glass, PPS, PEEK, Fluoroelastomer,
Acetal, Nitrile, EDPM
Sample Temperature Range
1°C – 50°C (34°F – 122°F)
Power Supply
100– 240 VAC, 47 - 63 Hz, 300 VA
Insulation Rating
Double Insulated, Pollution Degree 2, Overvoltage Category II
Environmental Conditions
The unit is intended for indoor use.
Up to 95% RH (non-condensing)
Regulatory Compliance
cLCus, Lead Free, conforms to CAN/CSA-C22.2 No. 61010-
And Certifications
1-12 (R2017)/UL 61010-1 (Ed. 3)
Shipping Weight
73 lbs
*Operating temperatures above 38°C may require the instrument to be primed before
readings or increase frequency of readings.
Viton® is a trademark of The Chemours Company FC, LLC
Noryl® is a registered trademark of SABIC Global Technologies B.V.
2.3 Unpacking and Inspection of
the Instrument
Remove all items from packing carton and carefully inspect to ensure
that no visible damage has occurred during shipment and that all items
listed below have been received. If the items received do not match
the order, please contact your local distributor or the HF scientific
Customer Service department.
• CSM - Copper Silver Monitor
• Instruction Manual
• Chemical Reagents, shipped separately

2.4 Instrument Labels

The following labels should be applied to the outside of the enclosure.
Purpose
Serial number, manufacturer,
part number, power rating,
UL marking
Informational
2.5 Display
1 - Home Button
2 - Configuration Menu Access
3 - Service Menu Access
4 - Date/Time of last Silver
reading
5 - Date/Time of last Copper
reading
6 - Last Silver reading
7 - Last Copper reading
8 - Cycles through current instrument status, warnings (yellow), errors
®
,
(red). If instrument is operating as normal, background will be green.

3.0 Installation

3.1 Site Selection and Mounting

The instrument is designed for wall mounting. The instrument should
be mounted at eye level and be easily accessible for operation and
service. Consideration must be made for the plumbing and electrical
conduit connections. Recommended mounting screws are M8 (
fasteners for the instrument enclosure.
It is critical that the instrument be mounted as close as possible
to the sampling point to ensure a quick response time (within 2
to 3 meters (6 to 10 feet) of the sampling point).
242mm
Air Vent
4
Location
Label
Side of Unit
Front of Unit
1
4
5
8
519mm
254mm
660mm
620mm
Power, Signal
Mounting Tabs
Sample In
Atmospheric Drain
Emergency Drain
2
3
6
7
")
5
16

Advertisement

Table of Contents
loading

This manual is also suitable for:

28062

Table of Contents