If some of the procedures do not suit your needs, please consult Pall or your local distributor before finalizing your system. Use of this product in a manner other than in accordance with Pall’s current recommendations may lead to injury or product loss.
Page 4
CAUTION! Avoid the use of sharp blades or pointed instruments that could damage the Basic/ Advanced Controlbox. Do not open the packaging by forcing any of the system components through the sealed end because this can generate particulate contaminants. Ensure that the system selected is suitable for the application.
On the top side you can read the values from the display. In case of the basic controlbox the temperature display is the only display. In the advanced one you can also read out pH and conductivity on a second display. Figure 4.2 : Basic Controlbox in UL www.pall.com/biotech...
Page 6
Figure 4.3 :Top view of Basic Controlbox (left) and Advanced Controlbox (right) Under the plug socket the cables for sensors, connection cables to LevMixer/Magnetic drive unit and weighing is located. As well you can find 2 Ethernet cable connections. Figure 4.4 : Right side of the cabinet plug socket for power and 2 Ethernet cable connections...
Figure 5.2 Main switches (left UL, right CE) WARNING! Fatal electric shock! Some spots can be live even when the main switch is switched off (refer to the wiring diagram). Only trained personnel may work on the electrical installations of the system. www.pall.com/biotech...
Page 8
MEASUREMENT CABLES-OVERVIEW In the following figure the different cables for the connection to the drive unit or for measurements via sensors is shown. Figure 5.3: Overview Advanced Controlbox cables 1- STATUS cable LevMixer/Magnetic Mixer 2- Conductivity cable 3- Agitator speed cable only for LevMixer drive unit 4- pH cable 5- Temperature cable 6- Weighing cable to connect with counterpart on the tank with indicator...
Page 9
Used for supplying power to the PLC control and sensors. It is a hardwired one for the UL versions. For the CE version it is a plug and play system to allow the Contolboxes a cable plus plug for a specified country ( refer to section 8). www.pall.com/biotech...
Page 10
AGITITOR SPEED Measurement of Agitator speed via standard cable and plug and play principle just for the LevMixer drive unit applicable (yellow cable). Refer to Figure 5.6 below. STATUS LevMixer and Magnetic Mixer Motor start /stop, remote or local status and alarm status is read by a grey cable connected via plug and play to the LevMixer or Magnetic Mixer.
Controlbox power socket. Connect the Ethernet cable Plug the Ethernet cable in the Ethernet socket for: o Service connection to OPC Interface o Connecting the system to an external PLC connection e.g MVP • Data transfer www.pall.com/biotech...
Page 12
Connect the sensors with the cables and install the sensors in the bag of the tank. Connect the weighing cable of the Basic/Advanced Controlbox (1) with the indicator cable of the weighing (2) equipped with a safety cap. Plug in the plug in the power socket of the Basic/Advanced Controlbox.
Page 13
Running the system. On the Basic/Advanced Controlbox screen of the display you can see the values. After your measurement is over, you reinstall your probes from the Mixer bag, disconnect the weighing cable and switch the main power of. www.pall.com/biotech...
6. Specifications Schematic System Overview On the following schematic the dependences are shown. It is divided in the end customer part, the Controlbox section and the Pall mixer tank section as well as the Pall drive unit and MVP section.
However as the legislation and facilities vary throughout the member states, please contact your local Pall sales office or distributor to discuss the available options for correctly disposing of this product.
The frame, stainless steel components and UCI can be cleaned using a soft cloth and a neutral detergent or alcohol solution. Do not use abrasive cleaning agents. Manufacturer Details The Basic/Advanced Controlbox is manufactured by: Pall GmbH Philipp-Reis-Str. 6 63303 Dreieich Germany...
Page 20
Canada, China, France, Germany, India, Indonesia, Ireland, Italy, Japan, Korea, Malaysia, New Zealand, Norway, +1.516.484.5400 phone Philippines, Poland, Russia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom, and Vietnam. Distributors in all major industrial areas of the world. To locate the Pall office or distributor nearest you, visit www.pall.com/contact. European Headquarters Fribourg, Switzerland The information provided in this literature was reviewed for accuracy at the time of publication.
Need help?
Do you have a question about the USD3295 and is the answer not in the manual?
Questions and answers