Safety; Instructions For Safe Operations; Warning; Attention - Contec CMS-50F User Instructions

Hide thumbs Also See for CMS-50F:
Table of Contents

Advertisement

1 Safety

1.1 Instructions for Safe Operations

Check the main unit and all accessories periodically to make sure that there is no visible
damage that may affect patient's safety and monitoring performance about cables and transducers. It
is recommended that the device should be inspected once a week at least. When there is obvious
damage, stop using the device.
Necessary maintenance must be performed by qualified service engineers ONLY. Users are not
permitted to maintain it by themselves.
The oximeter cannot be used together with devices not specified in User's Manual. Only the
accessory that appointed or recommendatory by manufacture can be used with this device.
This product is calibrated before leaving factory.

1.2 Warning

Explosive hazard—DO NOT use the oximeter in environment with inflammable gas such as
some ignitable anesthetic agents.
DO NOT use the oximeter while the testee measured by MRI and CT.
Please do not break the wristband, for fear it becomes out of use, or the unexpected drop of
the device which is due to the looseness of the wristband in the process of using. Users who are
allergic to the wristband is not recommended to use it.
The person who is allergic to rubber can not use this device.
The disposal of scrap instrument and its accessories and packings (including battery, plastic
bags, foams and paper boxes) should follow the local laws and regulations.
Please check the packing before use to make sure the device and accessories are totally in
accordance with the packing list, or else the device may have the possibility of working abnormally.
Please choose the accessories and probe which are approved or manufactured by the
manufacturer, or else it may damage the device.
Please choose the battery chargers which should be ensured compliance with the requirements
of IEC 60601-1, or else it may damage the device.
Please don't use the device in the course of charging.
The device can only be matched with the compatible probe.
Please don't measure this device with functional tester for the device's related information.

1.3 Attention

Keep the oximeter away from dust, vibration, corrosive substances, explosive materials, high
temperature and moisture.
If the oximeter gets wet, please stop operating it.
When it is carried from cold environment to warm or humid environment, please do not use it
immediately.
DO NOT operate keys on front panel with sharp materials.
High temperature or high pressure steam disinfection of the oximeter is not permitted. Refer to
User Manual in the relative chapter (7.1) for instructions of cleaning and disinfection.
Do not have the oximeter immerged in liquid. When it needs cleaning, please wipe its surface
1